Axetris AG
LGD Compact Operating & Integration Instructions; Rev. D
38/46
0xFD
W
Laser temperature compensation executed
(system restarting)
n/a
If this occurs more than twice per year please call
customer service.
0xFE
W
Alarm 1 out of defined limit
Customer & application specific
parameter
Sensor is working correctly. This parameter can be
changed in the sensor parameters by the user.
0xFF
W
Alarm 2 out of defined limit
Customer & application specific
parameter
Sensor is working correctly. This parameter can be
changed in the sensor parameters by the user.
0xEE
W
Sensor Line-Locking not secured
Contact the Axetris service if the
problem persists
Reference for line locking is below track-limit
Baseline noise above defined track-limit
ANNEX
A.
Status, Warning and Error Code Listing for Evaluation
E
rr
o
r
n
u
m
b
er
E
rr
o
r
/
W
ar
n
in
g
MESSAGE IN GUI
Action / Solution
Detailed explanation
0x13
E
Unexpected Error (Watchdog) . Reboot
(15s)
Check communication.
Contact service if problem persists
Check command structure and sensor connectivity.
0xEF
E
End of Life
Soon, sensor will no longer work
The sensor will soon be out of service.
0xF0
E
Opticalhead's temperature out of range
Check for additional errors (0xF9)
Bring the ambient system temperature
within the specified range
Try to mount the system horizontally.
Check if ambient temperature is within specified range.
See also error code 0xF8 & 0xFC in the user manual. If
OK (temperature within range), there is a system
failure: service required.
0xF1
E
Laser driver failure. Sensor not measuring. Check for additional errors (0xFA)
Service required
Check if ambient temperature is within specified range.
See also error code 0xF8 & 0xFC in the user manual. If
OK (temperature within range), there is a system
failure: service required.
0xF2
E
Laser driver failure. Sensor not measuring. Check for other errors (0xFB)
Service required
Policy not yet defiend. Reset the sensor.
0xF3
E
Low power on the photodiode.
Check for condensation OR
Service required
Check that the conditions of use do not cause
condensation. If this is not the case, service is
required.
0xF4
W
Low power on the photodiode.
Cleaning of optics required.
OR see error 0xFA in Manual
Sensor is functionning but with compromised
performance due to misaligned optics, dirt or
condensation. Service required soon. See maintenance
manual.
0xF5
W
Power on photodiode too high.
Service required
Sensor functionallity compromised. Please call servive.
0xF6
W
Laser temperature compensation executed
(system restarting)
-
Aging of the laser is compensated until it reached EOL
state.
0xF7
E
Laser current compensation limit reached
Service required.
EOL is reached. Call service immediately.
0xF8
E
Cell Temperature of out safe range.
Sensor not measuring.
-
SENSOR MIGHT BE HOT: DO NOT TOUCH WITH
BARE HANDS
The temperature limits are set between
(H) Version: {-40°C..200°C}, (A) Version: {-
50°C..100°C}.
Check ambient temperature.
Ensure proper venting
Actively monitor ambient conditions
out of range --> switch off sensor AND gas flow
Avoid condensation
0xF9
E
System temperature out of range. Sensor
not measuring.
Bring the ambient system temperature
within the specified range
Check ambient temperature.
Ensure proper venting
Activly monitor ambient conditions
out of range --> switch off sensor AND gas flow
Avoid condensation
0xFA
E
Laser driver failure. Sensor not measuring. Service required
Check if ambient temperature is within specified range.
See also error code 0xF8 & 0xFC in the user manual. If
OK (temperature in range), there is a system failure:
service required.
0xFB
E
Laser driver failure. Sensor not measuring. Service required