If you have DAPI / Hoechst in your sample, turn on the Diode 405 laser.  Then 
turn on the white light laser (WLL).  
8. Now get back to the Acquisition side to set up the excitation / detection 
paradigm. 

 
 
 
 
 
 

 

If you have any saved settings, you can load them from the button (A).  
 
If you need to make a new scheme you can add new sequences form the (F) 
button.  Then type the colour you have in your sample in the search window (B) 
and use mouse cursor to drag the desired dye from the dye bar (C) to the 
location of the desired detector (D) or (E).  Adjust detection ranges if necessary.  
 