Shutdown and preparing the microscope for the next
user
1.
Save your data.
2.
Check who is coming next to the microscope and what lasers they need
3.
Check from IRIS if the next user is coming in two hours or less.
If the next user is coming in 2 hours or less, use protocol A, else use
protocol B
A.
Turn off the 405 laser from the laser configuration menu if you used it and
the next user does not need it.
Shut down the software.
Remove your sample and clean the objectives.
Log off from the computer.
Check from IRIS if the next user within 24 hours needs the microscope
heated. If you used heating, leave the heating on, adjusting temperature if
needed. If you did not use heating, turn the heating on, adjusting
temperature if needed.
If you used CO
2
turn it off.
disinfect all the surfaces you touched.
B.
Turn off both the lasers from the laser configuration menu.
Shut down the software.
Remove your sample and clean the objectives.
Shut down the microscope.
Shut down the fluorescent lamp.