47
Leica RM2235
2. Attach vertical column
•
Insert the cap screw (
b
) into the hole of the
baseplate from below. Place the lock washer
(
c
) on the cap screw from above.
•
Thread the silver vertical column (
2
) onto the
baseplate from above and tighten with the
Allen key size 8.
The lock washer must be positioned
between the baseplate and vertical
column to prevent unintentional turn-
ing of the column.
A
Extreme danger of tipping! After install-
ing the vertical column, immediately
place the microtome onto the baseplate
so that the front feet of the microtome are
positioned in the shallow recesses (A).
3. Attach horizontal arm
•
Slide the support ring (
e
) onto the vertical
column and position it so that the lock nut (
f
)
is facing to the back. Tighten the lock nut.
•
Slide the cross-member (
d
) onto the column.
Ensure that the lock screw (
g
) is facing to the
right of the baseplate. The horizontal arm must
be centered over the microscope.
•
Slide the horizontal arm (
3
), flat side facing the
lock screw (
g
), into the cross-member (
d
) and
tighten.
For more information on connecting
and using the microscope, magnifying
lens or cold-light source, please see
the appropriate manual.