3.3 Technical data
General
Approvals:
The instrument-specific marks are located on the
rear panel of the instrument next to the name plate.
Operating temperature range:
+10 °C to +35 °C
Temperature range during storage:
+5 °C to +55 °C
Relative humidity:
max. 80 % non-condensing
Humidity during storage:
< 80 %
Section thickness range:
1.0 - 60.0 µm
Section thickness settings:
from 1.0 - 10.0 µm in 1.0 µm increments
from 10.0 - 20.0 µm in 2.0 µm increments
from 20.0 - 60.0 µm in 5.0 µm increments
Specimen feed:
approx. 24 mm, ± 2 mm
Vertical stroke:
70 mm
Max. sectioning range without retraction: 69 mm (without specimen orientation at 1 µm)
Max. sectioning range with retraction:
62 mm
The specimen retraction can be turned
off manually:
200 µm
Dimensions and weight
Width (including handwheel):
413 mm
Depth (including waste tray):
618 mm
Height (total):
305 mm (with tray on the hood)
Working height (knife blade):
168 mm (measured from the table)
Weight (without accessories):
approx. 37 kg
Optional equipment and optional accessories
Object orientation (option)
horizontal:
8°
vertical:
8 °
Rotating range:
± 90°
Trimming increments:
10 µm, 50 µm
Repositioning of knife holder base
north-south:
± 25 mm