Optional accessories
•
Push the section waste tray (
18
) from the
front to the microtome base plate (
53
) until it
is held in place by the two magnets (
39
) (on
the front of the microtome base plate).
To remove the section waste tray, lift it
slightly and pull it off towards the back.
Fig. 42
6.4 Section waste tray
53
18
39
The backlighting is inserted at the
front into the one-piece knife holder
base.
•
Remove the two screws (
1
) using a slotted
screwdriver and then remove the cover plate (
2
).
•
Insert the backlighting (
3
) in the recess at
the rear of the knife holder base.
•
Insert the plug (
4
) for the backlighting into
the socket (
5
) of the microtome and connect
the plug of the power adapter to an AC pow-
er socket.
The backlighting illuminates once the micro-
tome is turned on with the mains switch.
NEVER use the backlighting illumina-
tion with the super mega-cassette
clamp!
Fig. 43
2
1
3
4
1
5
6.5 Backlighting