13
Leica MZ16 A – System operation
Calibrating the
reference distance
(CD)
Canceling the
calibration
Initialize the calibration and select the calibration mode see on
the left.
Display after confirming the selection: CD 1.418 in the example
(CD = process start)
Look into the eyepieces.
Focus the specimen micrometer.
Adjust the magnification with the handwheel so that the
diameter of the small circle (inside line) corresponds to the
distance of 1.40 on the specimen micrometer.
Confirm and display the calibration
2
×
CAL
(beep).
Display (in the example) A 50: A = Marking of the zoom posi-
tion used for the calibration
Press
CAL
key at any zoom position.
Display: C dist. or C magn. depending on the selected mode.
SEL
key
Display: CD XXX or CM XXX depending on the selected mode
Press
CAL
for 2 seconds (2 beeps)
Display: To indicate that the calibration was canceled,
the
A
in front of the magnification is deleted.