Starting up
Display
Factory settings
Keys
8
Leica MZ16 A – System operation
System operation
Attach the power supply to the supply.
When the power cable is removed from the supply, the selec-
tion under
OPT
, the calibration and the diameter of the refer-
ence circles are maintained. The zoom positions saved with the
manual control are canceled.
Display at power-on:
• init and software version (e.g. SW V1.3):
The unit is being initialized. (Beep)
• MZ16 A initialization concluded (2 beeps)
• Actual magnification
Objective
1.0
×
Eyepiece
10
×
Coaxial illumination
no
CAL
C dist. (reference distance)
REF
FOV (field of view diameter)
Acoustic signal
on
▲ ▼
Function 1: Motor zoom up/down:
Function 2: in the Selection Key menu
OPT
ic
• Select eyepiece and objective factors
• Coaxial illumination (yes/no)
• Acoustic signal (on/off)
SEL
ect
Confirm and display the selection
CAL
ibration
Calibrating the motor zoom
REF
erence
Display diameter of reference circle for measurements
SEL + CAL
(simultaneously)
Travel to the calibrated zoom position
SEL
(press for approx. 2 seconds)
Reset user settings to factory settings
Display: def. val.