Leica M Series User Manual
Photography & Video 83
Photo Tubes and C-mounts
Application
All Leica DFC cameras are equipped with a
standardized C-mount interface. In turn, the
C-mount adapter for the respective trinocular
tube is connected to this interface. This adapter
creates a solid mechanical connection between
the microscope and camera and ensures opti-
mum rendering of the microscopic image on
the image sensor of the camera.
Usually, the ideal is for the digital camera to
capture as much of the field of vision as possi-
ble, while excluding as much of the black edge
of the field of vision as possible. To do so, the
magnification factor of the C-mount adapter
must match the image format of the sensor as
closely as possible (see table).
If there is unwanted shading at the corners even
with a compatible C-mount adapter installed, it
can be corrected using the "Shading function"
of the camera software.
Alternatively, you can also use a C-mount
adapter with higher magnification. This primar-
ily avoids the critical border area of the field of
vision and concentrates on the center of the
field of vision.
Cameras from third-party suppliers
In addition to Leica DFC cameras with the stan-
dardized C-mount interfaces, you can connect
third-party cameras to the microscope using a
T2 bayonet adapter.
Stereo 2016/V.1
Name
MPixel Sensor Size
Pixel-Size Cooling Interface
C-Mount*
MC170 HD
5
CMOS-c 1/2.3“ 2.35 µm
No
USB2 + HD
0.4×-0.5×
MC190 HD
10 CMOS-c 1/2.3“ 1.67 µm
No
USB2 + HD
0.4×-0.5×
DMC2900
3
CMOS-c 1/2“
3.2 µm
No
USB3
0.5×
DFC295
3
CMOS-c 1/2“
3.2 µm
No
FWb
0.5×
DFC7000T
2.8 CCD-c 2/3“
4.54 µm
Yes
USB3
0.63×
DMC4500
5
CCD-c 2/3“
3.4 µm
No
USB3
0.63×
DFC450 C
5
CCD-c 2/3“
3.4 µm
Yes
FWb + Sync
0.5×
DFC495
8
CCD-c 2/3“
2.7 µm
Yes
FWb + Sync
0.63×
DFC550
12.5 CCD-c 2/3“
6.45 µm
Yes
FWb + Sync
0.63×
DFC3000 G
1.3 CCD-m 1/3“
3.75 µm
No
USB3 + Sync 0.35×-0.4×
DFC365 FX
1.4 CCD-m 2/3“
6.45 µm
Yes
FWb + Sync
0.7×
-c: color / -m: monochrome / * recommended
To do so, instead of the C-mount adapter,
simply use the corresponding SLR adapter
with T2 connection. However, these third-party
cameras are not integrated into the Leica Appli-
cation Suite and have to be operated using
the corresponding software from the camera
manufacturer.
The Leica digital cameras are detailed in a sepa-
rate user manual along with instructions for
their assembly and use.
Summary of Contents for M165 FC
Page 1: ...Leica M165 FC Leica M205 FA M205 FCA User Manual ...
Page 19: ...Leica M Series User Manual Assembly 19 Assembly ...
Page 64: ...Leica M Series User Manual Quick Start Guide 64 Quick Start Guide ...
Page 75: ...Leica M Series User Manual Eyepieces 75 Eyepieces ...
Page 81: ...Leica M Series User Manual Photography Video 81 Photography Video ...
Page 87: ...Leica M Series User Manual Microscope Carrier 87 Microscope Carrier ...
Page 91: ...Leica M Series User Manual Bases 91 Bases ...
Page 107: ...Leica M Series User Manual System Illumination 107 System Illumination ...
Page 116: ...Leica M Series User Manual Fluorescence 116 Fluorescence ...
Page 126: ...Leica M Series User Manual Accessories 126 Accessories ...
Page 131: ...Leica M Series User Manual Dimensional Drawings 131 Dimensional Drawings ...
Page 144: ...Leica M Series User Manual Specifications 144 Specifications ...
Page 156: ...Leica M Series User Manual Appendix 156 Appendix ...