Leica M Series User Manual
System Illumination 114
Leica EL6000 – Troubleshooting
Fault description
Possible faults
Remedy
Power indicator is not illuminated after being
switched on, instrument shows no function
No power supply voltage present; connecting
cable not plugged in correctly/defective
Check power supply voltage and connecting
cable
Fuses in instrument defective
Replace fuses, use permitted replacement
type only (see page 115)
Power switch is illuminated after being switched
on, instrument shows no function
Housing cover loose
or removed
Attach housing cover securely
Lamp worn out
Replace lamp
Target system receives too little light or none at all
Shutter closed or intensity switch at left limit
stop
Open shutter or rotate intensity switch
towards the right
Light guide not inserted into light guide
output properly
Check light guide
Light guide damaged/defective
Repair or replace
If these actions do not fix the problem, send the Leica EL6000 compact
light source to Leica service with an exact description of the malfunction.
Summary of Contents for M165 FC
Page 1: ...Leica M165 FC Leica M205 FA M205 FCA User Manual ...
Page 19: ...Leica M Series User Manual Assembly 19 Assembly ...
Page 64: ...Leica M Series User Manual Quick Start Guide 64 Quick Start Guide ...
Page 75: ...Leica M Series User Manual Eyepieces 75 Eyepieces ...
Page 81: ...Leica M Series User Manual Photography Video 81 Photography Video ...
Page 87: ...Leica M Series User Manual Microscope Carrier 87 Microscope Carrier ...
Page 91: ...Leica M Series User Manual Bases 91 Bases ...
Page 107: ...Leica M Series User Manual System Illumination 107 System Illumination ...
Page 116: ...Leica M Series User Manual Fluorescence 116 Fluorescence ...
Page 126: ...Leica M Series User Manual Accessories 126 Accessories ...
Page 131: ...Leica M Series User Manual Dimensional Drawings 131 Dimensional Drawings ...
Page 144: ...Leica M Series User Manual Specifications 144 Specifications ...
Page 156: ...Leica M Series User Manual Appendix 156 Appendix ...