Leica M Series User Manual
Eyepieces 78
All Leica eyepieces are also available with built-
in dioptric correction, allowing the microscope
to be used without glasses even by those with
vision problems. The correction comprises ±5
diopter settings.
Using the dioptric correction
1. Set the dioptric correction of both
eyepieces to the mid position ("0" diopter
settings).
2. While wearing your glasses, look through
the eyepieces and focus on the specimen.
3. Rotate both eyepieces to the maximum
value of "+5".
4. Hold one eye closed and rotate the other
eyepiece in the "-" direction until the speci-
men appears sharp.
5. Then, open the other eye and correct the
diopter settings until the image is uniformly
sharp.
Note that when using dioptric correction,
the advantage of parfocality is lost – thus
you have to manually refocus each time you
change the zoom level. To also use parfocality
with dioptric correction, refer to the instruc-
tions on page 79.
Dioptric Correction
Summary of Contents for M165 FC
Page 1: ...Leica M165 FC Leica M205 FA M205 FCA User Manual ...
Page 19: ...Leica M Series User Manual Assembly 19 Assembly ...
Page 64: ...Leica M Series User Manual Quick Start Guide 64 Quick Start Guide ...
Page 75: ...Leica M Series User Manual Eyepieces 75 Eyepieces ...
Page 81: ...Leica M Series User Manual Photography Video 81 Photography Video ...
Page 87: ...Leica M Series User Manual Microscope Carrier 87 Microscope Carrier ...
Page 91: ...Leica M Series User Manual Bases 91 Bases ...
Page 107: ...Leica M Series User Manual System Illumination 107 System Illumination ...
Page 116: ...Leica M Series User Manual Fluorescence 116 Fluorescence ...
Page 126: ...Leica M Series User Manual Accessories 126 Accessories ...
Page 131: ...Leica M Series User Manual Dimensional Drawings 131 Dimensional Drawings ...
Page 144: ...Leica M Series User Manual Specifications 144 Specifications ...
Page 156: ...Leica M Series User Manual Appendix 156 Appendix ...