HistoCore Water Bath
Setting up the instrument
4
4. Setting up the instrument
4.1 Installation site requirements
Danger
Hazardous substances (combustible or flammable) are stored near the instrument or filled in the water
tray.
People injury due to the explosion or ignition of the hazardous substances.
• Do not store combustible or flammable substances near the instrument.
• Use non-flammable liquid (preferably distilled water) only.
•
When installing the power cable, always be sure to route it so that it cannot contact the heated
surfaces of the instrument at any time.
•
The power socket to which the instrument is to be connected to has to be located near the instrument
and easily accessible.
• The power supply must be at a distance no greater than the length of the power cable – an extension
cable must not be used.
• The substrate must be largely vibration-free and have sufficient load capacity and rigidity for the weight
of the instrument.
• Avoid impacts, bright direct light, and excessive temperature fluctuations.
•
The instrument must be connected to a suitable power socket. Use only the provided power cable,
which is intended for the local power supply.
4.2 Standard delivery-packing list
Quantity
Part description
Order No.
1
HistoCore Water Bath
14 0607 020C0
1
basic instrument
14 0607 02000
1
Set of fuses
14 6000 05950
1
Instructions for Use (printed in English with additional languages on
data carrier 14 0607 80200)
14 0607 80001
If the supplied local power cord is defective or lost, please contact your local Leica representative.
Note
•
Please check all delivered parts against the packing list and against your order to verify whether
the delivery is complete. Should you find any discrepancies, please contact your Leica Biosystems
sales office without delay.