18
Version 1.0, Revision A
Setting up the instrument
4
Connecting the slide dryer(s) to the water bath
1. Remove the corresponding covers
on the bottoms of the wather bath and the
slide dryer by loosening screws
Fig. 13
1
2
2. Pull out the cables from two devices. Connect the cables together
. You will hear a
click sound and feel the connectors fitting into each other. If two slide dryers are to be installed
in parallel at the same side of the water bath, use the extension cable to connect the water bath
and the outer slide dryer
Fig. 14
1
2
Setting up the power supply
1. Before connecting the power cable, make sure that the main switch
on the front
of the instrument is at "O" (OFF) position.
Fig. 15
1