![Leica DMI6000 B Operating Manual Download Page 49](http://html1.mh-extra.com/html/leica/dmi6000-b/dmi6000-b_operating-manual_1873312049.webp)
49
6. Assembly
Now open the drawer several mm further until it
clicks into its end position. The disk will no long-
er turn in this position.
You can now insert a filter block. Proceed as fol-
lows:
• With the holder facing you squarely, insert the
filter or reflector cubes into the holder in
accordance with the included identification
sheet.
• The fluorescence cubes are suitable for both
upright and inverted microscopes. When
using them with inverted microscopes, insert
them so that the writing is upside down along
the lower edge.
To do so, place the filter or reflector cube on
the
left
side and press it to the
right
into the
mounting (Fig. 68).
• Make sure that the cube is correctly seated. A
loose cube can be destroyed or block the disk.
• For the next cube, close the drawer to the
point that the disk is once again free to turn.
Once you have reached the next position,
open the drawer fully once again. Continue in
this way for all of the cubes.
• Once all filter and reflector cubes have been
inserted, close the drawer and replace the
analyzer or cap.
Replacing cubes with the instrument switched
on:
• Remove the analyzer or the cap of the
analyzer slot.
• Press and hold the SET button on the right
side of the stand and at the same time, press
the button on the front panel for the cube you
would like to insert or replace.
• The filter changer will then rotate to the
correct position to insert or replace the cube
when you open the drawer by pressing the
white button on the right side of the stand.
The following message will appear in the top
line of the LeicaScreen.
Load.
To insert the cubes, proceed exactly as de-
scribed above.