6. Assembly
6.8 Installation of Filters in the Illumination Arm
The Leica DMI 6000 B is generally equipped with
a holder for two 40mm diameter filters. The fil-
ters are installed at the factory. To change filters
yourself, proceed as follows:
Release the screw (46.1) and remove the
cover.
Place the filter in the holder.
Place the cover on the transmitted light
illumination carrier and fasten with the
locking screw.
• Mark the lever with the provided adhesive
labels.
Fig. 46
Unscrewing the filter holder cover and inserting
filters in the transmitted light illumination arm
1
Screw
6.9 Installing the Transmitted Light Lamp
Housing
• Place the lamp housing in the transmitted light
lamp housing mount (fig. 47) and fasten it with
the clamping screw on the side.
• Thread the cable through the transmitted light
illumination arm (Fig. 48).
• Connect the lamp housing cable to the power
supply for transmitted light on the Leica
CTR6000 electronics box (Fig. 49.1).
For instructions on changing the lamp, please
see Chapter 6.10.
These instructions also apply to installing an Hg
lamp on the transmitted light axis. For descriptions
of the lamp housings and replacement of the
burner, please see Chapter 6.12,
→
p. 44ff.
1
Fig. 49
Connecting the lamp housing to the
Leica CTR6000 electronics box
Fig. 48
Lamp housing cabling (cable duct)
Fig. 47
Mounting the lamp housing on the
transmitted light illumination arm
1