26
6. Assembly
Fig. 17
Lamp mount with 12 V 100 W halogen bulb
Inserting the 12 V 100W halogen bulb into the
106/106 z lamp housing
• Unscrew the fastening screws of the cover
and flip the cover up (16.1).
• Unscrew the fastening screws of the lamp
mount (16.8) and pull out the mount (Fig. 17).
• Insert the lamp with the dust cover straight
into the socket until it stops.
Caution!
Do not remove the lamp’s dust cover until af-
ter you have installed the lamp. Avoid finger-
prints on the lamp.
• Remove the dust cover.
• Insert the lamp mount, with the burner in-
stalled, into the lamp housing and tighten it
with the screws (16.8).
• Close the lamp housing and retighten the fas-
tening screws.
• Place the lamp housing in the incident light
lamp housing receptacle (18.1) and fasten it
with the clamping screw on the side.
• Connect the lamp housing to the power supply
for incident light (symbol:
) (18.4).
Fig. 18
Rear view of stand
1
Incident light lamp housing receptacle
2
Transmitted light lamp housing receptacle
3
12 V 100 W connection for transmitted light (symbol:
)
4
12 V 100 W connection for incident light (symbol:
)
1
2
3 4