29
6. Assembly
Fig. 22
Rear panel of the ebq 100 supply unit
1
Lamp connection
• Insert the lamp mount, with the burner in-
stalled, into the lamp housing and tighten it
with the screws (20.8).
• Close the lamp housing and retighten the fas-
tening screws.
• Place the lamp housing in the incident light
lamp housing receptacle (21.1) and fasten it
with the clamping screw on the side.
• Connect the lamp housing to the external
power supply (22.1).
1
Fig. 20
106/106 z lamp housing (on the side, open)
1
Cover raised
2
Collector
3
12 V 100 W lamp or
gas discharge lamp in mount
4
Reflector (mirror)
5, 6, 7
Adjusting screw for x-y reflector
8
Fastening screw for lamp mount
9
Socket for contact plug
1
2
3
8
9
8
4
5
6
7
Fig. 21
Rear view of the stand
1
Incident light lamp housing receptacle
2
Transmitted light lamp housing receptacle
3
12 V 100 W connection for transmitted light (symbol:
)
4
12 V 100 W connection for incident light (symbol:
)
1
2
3 4