24
6. Assembly
• Insert the new 12 V 100 W lamp (13.1) with the
dust cover straight into the socket until it
stops. Be sure that the lamp is inserted
straight.
• Remove the lamp’s dust cover.
Caution!
Do not remove the lamp’s dust cover until af-
ter you have installed the lamp. Avoid finger-
prints on the lamp.
• Replace the housing and fasten it in place us-
ing the fastening screw.
Fig. 12
Lamp housing 107/2
Releasing the
fastening screw
1
2
Fig. 13
Lamp housing 107/2
opened
1
Mount with
halogen bulb
2
Collector
Fig. 14
Rear view of the stand
1
Incident light lamp housing receptacle
2
Transmitted light lamp housing receptacle
3
12 V 100 W connection for transmitted light (symbol:
)
4
12 V 100 W connection for incident light (symbol:
)
• Place the lamp housing in the transmitted
light lamp housing receptacle (14.2) and fas-
ten it with the clamping screw on the side.
• Connect the lamp housing to the power supply
for transmitted light (symbol:
) (14.3).
1
2
3 4