ZS1000 High Impedance Active Probe
A–2
ZS1000-OM-E Rev C
EQUIPMENT USED:
1The function generator used in this Performance Verification Procedure is used for mak-
ing relative measurements. The output of the generator is measured with a DMM or
oscilloscope in this procedure. Thus, the generator is not required to be calibrated.
ZS1000 TEST RECORD
Step
Description
Intermediate data
Test Result
Output Zero Voltage
A-5
Output Zero Voltage
(Test limit
≤
±800 µV)_____________ V
LF Attenuation Accuracy
B-7 Generator Output Voltage
______________ V
B-8 Expected Output Voltage, top range
______________ V
B-10 Measured Output Voltage, top range
______________ V
B-12
Gain Error
, top range (Test limit
≤
±1.0%) ______________%
B-17 Generator Output Voltage
______________ V
B-19 Expected Output Voltage, mid range
______________ V
B-21 Measured Output Voltage, mid range
______________ V
B-23
Gain Error
, mid range (Test limit
≤
±1.0%) ______________%
###
MODEL
SERIAL
NUMBER
CALIBRATION
DUE DATE
OSCILLOSCOPE
DIGITAL MULTIMETER
FUNCTION
GENERATOR
1
N/A