ZS1000 High Impedance Active Probe
5–4
ZS1000-OM-E Rev C
where:
Z
probe
is the probe’s input impedance and
Z
source
is the source impedance
As an example:
At 750 MHz, where the probe input impedance has reduced to 236
Ω,
and a source resistance of 250
Ω
the probe output amplitude is reduced
to:
###
V
out
236
236 250
+
------------------------
0.49
V
in
×
=
=