ZS1000 High Impedance Active Probe
8–4
ZS1000-OM-E Rev C
7. Disconnect the probe from the ProBus extender and re-install the
circuit board into the probe case, being careful to align the ProBus
interface connector with the opening on the other end of the probe.
Figure 8-2. Adjustment Locations
C. Verify Calibration
Repeat the Performance Verification procedure to ensure compliance
with the warranted specifications.
Apply a calibration sticker, if required, in accordance with your quality
control procedures.
###