2. ETHERNET
26
Command Parameter1
Parameter2
Description
10 to 200
Sets the upper limit (%) of the timing
jitter for HD.
SYS:UNIT:ERROR:HD_TIMING_JIT:MAX 1,2,3,4
?
Queries the upper limit of the timing
jitter for HD.
ON
Detects the error of current jitter value
for HD.
OFF
Does not detect the error of current jitter
value for HD.
SYS:UNIT:ERROR:HD_CURRENT_JIT 1,2,3,4
?
Queries the presence of error detection
of current jitter value for HD.
10 to 200
Sets the upper limit (%) of the current
jitter for HD.
SYS:UNIT:ERROR:HD_CURRENT_JIT:MAX 1,2,3,4
?
Queries the upper limit of the current
jitter for HD.
ON
Enables the amplitude error detection
for SD.
OFF
Disables the amplitude error detection
for SD.
SYS:UNIT:ERROR:SD_AMP 1,2,3,4
?
Queries the presence of the amplitude
error detection for SD.
80 to 140
Sets the upper limit (%) of the amplitude
error for SD.
SYS:UNIT:ERROR:SD_AMP:UPPER 1,2,3,4
?
Queries the upper limit of the amplitude
error for SD.
40 to 100
Sets the lower limit (%) of the amplitude
error for SD.
SYS:UNIT:ERROR:SD_AMP:LOWER 1,2,3,4
?
Queries the lower limit of the amplitude
error for SD.
ON
Enables the error detection of the rise
time for SD.
OFF
Disables the error detection of the rise
time for SD.
SYS:UNIT:ERROR:SD_RISE 1,2,3,4
?
Queries the presence of the error
detection of the rise time for SD.
40 to 140
Sets the upper limit (%) of the rise time
for SD.
SYS:UNIT:ERROR:SD_RISE:MAX 1,2,3,4
?
Queries the upper limit of the rise time
for SD.
ON
Enables the error detection of the fall
time for SD.
OFF
Disables the error detection of the fall
time for SD.
SYS:UNIT:ERROR:SD_FALL 1,2,3,4
?
Queries the presence of the error
detection of the fall time for SD.
40 to 140
Sets the upper limit (%) of the fall time
for SD.
SYS:UNIT:ERROR:SD_FALL:MAX 1,2,3,4
?
Queries the upper limit of the fall time
for SD.