RECOMMENDATIONS FOR CORRECT INSTALLATION OF THE LOAD CELLS
SIZING OF LOAD CELLS CAPACITY
For safety reasons, in case of static weighing, it is advisable to use the load cells at a maximum of
70-80% of its nominal capacity (assuming that the load is uniformly distributed over the entire
weighed structure); depending on the handling mode of the load to weigh, consider to further reduce
the % of load with respect to the nominal capacity (ex.: forklifts handling, bridge cranes, etc.).
In case of weighing with dynamic loads, the installer has to estimate the thrust speed, the
acceleration, the frequency, etc.
INSTALLING LOAD CELLS
The load cells must be placed on rigid, stable in-line structures; it is important to use the mounting
modules for load cells to compensate for misalignment of the support surfaces.
PROTECTION OF THE CELL CABLE
Use water-proof sheaths and joints in order to protect the cables of the cells.
MECHANICAL RESTRAINTS (pipes, etc.)
When pipes are present, we recommend the use of hoses and flexible couplings with open
mouthpieces with rubber protection; in case of hard pipes, place the pipe support or anchor bracket as
far as possible from the weighed structure (at a distance at least 40 times the diameter of the pipe).
WELDING
Avoid welding with the load cells already installed. If this cannot be avoided, place the welder ground
clamp close to the required welding point to prevent sending current through the load cell body.
WINDY CONDITIONS - KNOCKS
–
VIBRATIONS
The use of weigh modules is strongly recommended for all load cells to compensate for
misalignment of the support surfaces. The system designer must ensure that the plant is protected
against lateral shifting and tipping relating to: shocks and vibration; windy conditions; seismic
conditions in the installation setting; stability of the support structure.