- 16 -
By confirming with ENTER the setpoint operation can be set to the value 0:
-
: relay switching will not occur if the setpoint value is 0.
-
:
-
setpoint = 0 and switching =
: relay switching occurs when the weight is 0; the relay
will switch again when the weight is different from zero, taking hysteresis into account (both
for positive and for negative weights).
-
setpoint = 0 and switching =
: relay switching occurs for a weight higher than or equal to
0, the relay will switch again for values below 0, taking hysteresis into account.
-
setpoint = 0 and switching =
: relay switching occurs for a weight lower than or equal to
0, the relay will switch again for values above 0, taking hysteresis into account.
INPUTS
Default: input 1 =
input 2 =
Possible operation modes:
-
(NET/GROSS): by closing this input for no more than one second, it’s making an
operation of SEMI-AUTOMATIC TARE and the display will show the net weight. To display the
gross weight again, hold the NET/GROSS input closed for 3 seconds.
Function not available if net functions are disabled (see parameter
in section
INFO MENU
).
-
: by closing the input for no more than one second, the weight is set to zero (see section
WEIGHT ZERO-SETTING FOR SMALL VARIATIONS (SEMI-AUTOMATIC ZERO)
).
-
: closing the input no operation is performed, the input status may however be read remotely
by way of the communication protocol.
-
: closing the input for max one second the weight is transmitted over the serial
connection according to the fast continuous transmission protocol only once (
only if
is
set in the item
).
-
:
when the input is closed the data are sent for printing if in the communication protocol
of either serial port the parameter
is set. If the alibi memory is active, data storage is
carried out too.
-
:
replicates the function of IN.
-
:
replicates the function of OUT.