Installation and Getting Started
User’s Manual Switchos
D850R3
2-3
2.3
Installing the Switchos with the UltiMate Capillary HPLC
System and the FAMOS Microautosampler
When the Switchos is used as a component in the LC Packings UltiMate Capillary
HPLC system, all instruments are controlled by the CHROMELEON software. The
loading pump will be included in the instrument network and the switching
valves are controlled by a COM port of the PC.
2.3.1
Electrical Connections
All electrical connections are made on the rear panel of the instrument
(FIGURE 2-1).
SOLVENT SELECTION
Button
REMOTE/LOCAL
Switch
COMMUNICATION
Connector
INPUTS
Connector
Power Outlet to
Loading Pump
Main Power Switch
Main Power Inlet
Fuse
He Inlet
Power Inlet
RS-232-1
Connector
RS-232-2
Connector
EVENTS
Connector
REMOTE
Connector
Loading Pump
Power Switch
FIGURE 2-1 Rear Panel of the Switchos Advanced Microcolumn Switching Unit
CAUTION
Caution: Avoid touching the electrical contacts on the terminal strips.
Electrostatic discharges could damage internal components.
The manufacturer will not accept any liability for damages directly or indirectly
caused by connecting the Switchos Advanced Microcolumn Switching Unit to
instruments which do not meet relevant safety standards.
Summary of Contents for SWITCHOS II
Page 6: ...Table of Contents iv D850R3 User s Manual Switchos II This page intentionally left blank...
Page 14: ...CE Declaration xii D850R3 User s Manual Switchos II This page intentionally left blank...
Page 22: ...Introduction 1 8 D850R3 User s Manual Switchos II This page intentionally left blank...
Page 80: ...The Switching Valves A 4 D850R3 User s Manual Switchos II This page intentionally left blank...
Page 82: ...Network Identification B 2 D850R3 User s Manual Switchos II This page intentionally left blank...