12
Modifying the time internal parameters
In order to enter the temperature internal parameter setting mode, the user must introduce the
password. To do that, follow these steps:
1.
Press
for 3 seconds.
-
The screen will display
Lc
-
The speed area of the screen will display the password value
2.
Use
to choose the required password value. See password values below.
3.
Press
to enter the value.
If the password is
incorrect
, the display will return to the home screen.
If the password is
correct
, the display will show the temperature internal parameter settings mode.
Then:
1.
Press
to modify each parameter.
2.
Use
to choose desired value.
3.
Press
for 3 seconds to return to the home screen.
If no key is pressed for 1 minute, the display will return to the home screen.
5.2. Internal parameter table
INDICATOR
NAME
FUNCTION DESCRIPTION
FACTORY VALUE
Lc
Password
The password to modify the parameters in this
table is
Lc = 3
0
ndT
Timing mode
selection
0:
when the timer goes off, stop the speed only,
not the temperature.
1:
when the timer goes off, stop the speed AND
the temperature.
2:
The timer starts to count after reaching the
temperature set value. When the timer goes off,
stop the speed only, not the temperature.
3:
The timer starts to count after reaching the
temperature set value. When the timer goes off,
stop the speed AND the temperature.
Note: This parameter cannot be modified during
operation. If you do that, the timer will restart.
(0
~
3)
1
Hn
Total time
units
0:
Time is measured in minutes
–
m
is displayed.
1:
Time is measured in hours
–
h
is displayed.
Note: This parameter cannot be modified during
operation.
(0
~
1)
0
SPd
Constant
temperature
deviation
IF
THEN the equipment enters the constant
temperature state.
(0.1
~
100.0
℃
)
0.5
temperature
measurement
temperature
set value
SPd
≥
-
Summary of Contents for WB-S