STANDARD USER MANUAL n° #0020-3
SMU#0020-03
Page 11/25
Warnings:
1)
Oxygen outlet must be maintained at atmospheric pressure.
2)
If H
2
outlet under pressure is suddenly opened, in certain circumstances H
2
could
be mixed with water. Switch OFF device before disconnecting H
2
Outlet.
CAUTION
: your HYDROXYCHROM generator has been tested for several hours at the factory and all its
tubes have been cleaned of ambient air contaminants. After a break of several days to several weeks of
operation, Air ambient has slightly dirt the circuit of the instrument. The walls of the tubes need again to be
cleaned. Before connecting to the consumer, please let your H2-generator running during a couple of hours
to the atmosphere.
4.4.2.
E
LECTRIC CONNECTIONS
Mains:
Connect the HYDROXYCHROM generator with the electrical cable provided; if it was not possible,
verify that the cable use has a sufficient section and has a ground wire (3X1 mm
2
minimum). Make sure
laboratory differential circuit breaker can absorb an inrush current of at least 6A without switching off.
4.5.
F
ACTORY
S
ETTINGS
During one of the last phases of QC, your HYDROXYCHROM has been programmed with a set of values
called “factory settings”. Those settings will help you to start the generator without troubles:
Set H
2
Pressure: 2000mbar
Timeout Pressure Alarm: 0 second (timeout disabled)
Functioning Mode: Continuous on
At this step, the HYDROXYCHROM is ready for startup.
Switch power ON
After a few seconds due to generator initialization, the water pump1 sucks ZEROWATER and fill up
the internal tank
When upper water level is reached, current is applied to the hydrogen cell. Electrolysis process is
initiated and some H
2
is available at the H
2
outlet.
The H
2
generator’s configuration could be modified according to your needs via its USB connection and a PC
with
HYDROXYCHROM Viewer
software.