Warnings: 

1) 

Oxygen outlet must be maintained at atmospheric pressure. 

2) 

If H

2

 outlet under pressure is suddenly opened, in certain circumstances H

2

 could 

be mixed with water. Switch OFF device before disconnecting H

2

 Outlet. 

 

CAUTION

:  your  HYDROXYCHROM  generator  has  been  tested  for  several  hours  at  the  factory  and  all  its 

tubes  have  been  cleaned of  ambient  air  contaminants.  After  a  break  of  several  days to  several weeks of 
operation, Air ambient has slightly dirt the circuit of the instrument. The walls of the tubes need again to be 
cleaned. Before connecting to the consumer, please let your H2-generator running during a couple of hours 
to the atmosphere. 

 

4.4.2.

 

E

LECTRIC CONNECTIONS

 

Mains:

  Connect  the  HYDROXYCHROM  generator with the electrical cable provided;  if it was not possible, 

verify  that  the  cable  use  has  a  sufficient  section  and  has  a  ground  wire  (3X1 mm

2

  minimum). Make  sure 

laboratory differential circuit breaker can absorb an inrush current of at least 6A without switching off. 

 

4.5.

 

F

ACTORY 

S

ETTINGS

 

During  one  of  the  last  phases  of  QC,  your  HYDROXYCHROM  has  been  programmed  with  a  set  of  values 
called “factory settings”. Those settings will help you to start the generator without troubles:  

 

Set H

2

 Pressure: 2000mbar 

 

Timeout Pressure Alarm: 0 second (timeout disabled) 

 

Functioning Mode: Continuous on 

 

At this step, the HYDROXYCHROM is ready for startup. 

 

Switch power ON 

 

After a few seconds due to generator initialization, the water pump1 sucks ZEROWATER and fill up 

the internal tank  

 

When upper water level is reached, current is applied to the hydrogen cell. Electrolysis process is 

initiated and some H

2

 is available at the H

2

 outlet. 

 

The H

2

 generator’s configuration could be modified according to your needs via its USB connection and a PC 

with 

HYDROXYCHROM Viewer

 software. 

 

 

Summary of Contents for HYDROXYCHROM

Page 1: ...e Application date Modified chapters 0 Creation 24 10 2012 All 1 Revision 25 10 2012 Ch 3 4 2 Revision 22 04 2015 Ch 4 6 2 3 Revision 14 12 2018 Ch 3 1 4 3 Editor S GUTIERREZ JP AMIET K TRAULLE Checke...

Page 2: ...IONING PRINCIPLE 8 CHAPTER 4 INSTALLATION AND OPERATION 9 4 1 RECEIPT OF INSTRUMENT AND CHECK 9 4 2 DELIVERY CONTENT 9 4 3 GENERATOR INSTALLATION 9 4 4 CONNECTIONS 10 4 4 1 Fluidic connections 10 4 4...

Page 3: ...the delivery date The replacement of the defective components will be free except the transport or travel fees that will be invoiced according the current tariffs CHROMATO SUD accepts no liability fo...

Page 4: ...operations alarms as well as troubleshooting It is applicable to the below generation models and software HYDROXYCHROM Viewer Version 1 0 produced since October 2012 HYDROXYCHROM 100 HYDROXYCHROM 160...

Page 5: ...cause 2 2 W E E E PRODUCT RECYCLING DECLARATION In agreement with the European EC 2002 96 directive on electrical and electronic equipment recycling this product may not be disposed in the garbage Fo...

Page 6: ...d water TOC free Conductivity 0 20 S cm Water capacity 5L tank outside 0 4L tank inside Water consumption 5L water generates about 6000L Hydrogen Safety Low H2 stored volume over pressure valve intern...

Page 7: ...ear panel left side has Mains plug with fuse and switch DB9 interface connectors for RS485 remote control USB dry relay The rear panel in the center has ZEROWATER input from external tank fitting for...

Page 8: ...YDROXYCHROM 100 and HYDROXYCHROM 160 are equipped with a single layer H2 cell H2 GENERATION MODULE O2 Outlet Water Inlet H2 Outlet External 5L Water Tank ZERO WATER box Water pump 1 Static dryer 24 H2...

Page 9: ...charged to the customer It is well recommended to keep the wood box for future shipments 4 2 DELIVERY CONTENT Quantity Description 1 Hydrogen Generator HYDROXYCHROM with its factory communication opt...

Page 10: ...at a maximum distance of 1 2 m from the generator It can nevertheless be placed at the same level or above the instrument The maximum difference in height between the generator and the ZEROWATER tank...

Page 11: ...it was not possible verify that the cable use has a sufficient section and has a ground wire 3X1 mm2 minimum Make sure laboratory differential circuit breaker can absorb an inrush current of at least...

Page 12: ...provides User Interface to control Gas Generator Network through USB interface 4 6 1 SPECIFICATIONS Requirements PC or Laptop under WinXP SP2 or higher Win Vista Win 7 At least 5Mb of free space on th...

Page 13: ...you must log the computer in administrator Remark If you want work with HydroxychromViewer in Russian start by installing the Cyrillic code page procedure below and then come back here only available...

Page 14: ...STANDARD USER MANUAL n 0020 3 SMU 0020 03 Page 14 25 Step 3 Choose destination folder or use default Step 4 Complete installation...

Page 15: ...chrom Viewer V1 0 when the installation is complete you have to do a commit cf procedure below If you just installed Hydroxychrom Viewer V2 0 when the installation is complete this window will appear...

Page 16: ...STANDARD USER MANUAL n 0020 3 SMU 0020 03 Page 16 25 CYRILLIC CODE PAGE INSTALATION Step 1 Run install_cyrillic_codepage exe Step 2 Select your preferred language for this installation...

Page 17: ...tallation When the installation is complete this window will appear clic on yes the computer will restart automatically saving the modification you just made Remark After the computer restarted you ca...

Page 18: ...EDURE FOR COMPUTER UNDER EMBEDDED WINDOWS Open the program named command prompt in the toollbar In this program write commitcf as shown in the picture below and press enter The program asks you to res...

Page 19: ...STANDARD USER MANUAL n 0020 3 SMU 0020 03 Page 19 25 CHANGING LANGUAGE Step 1 Open Application partition Step 2 Open Hydroxychrom folder Step 3 Open Hydroxychrom_v2 ini file with Notepad...

Page 20: ...ep 4 Modify the second line of the code LanguageFile Hydroxychrom_UK lng for English language LanguageFile Hydroxychrom_FR lng for French language LanguageFile Hydroxychrom_RU lng for Russian language...

Page 21: ...R MANUAL n 0020 3 SMU 0020 03 Page 21 25 And open it again USAGE Step 1 Connect USB Step 2 Start HYDROXYCHROM Viewer Step 3 Adjust the COM port number that is associated with USB port Step 4 Left Clic...

Page 22: ...f following Off No gas generation On Gas generation is running but will be stopped after power off Continuous On Gas generation is running but will be automatically started every time after power on R...

Page 23: ...ure does not match the Set Pressure Timeout Alarm is time in seconds given to the instrument after start to reach the Set Pressure IMPORTANT NOTE For security reasons the Instrument stops generation w...

Page 24: ...ter should be exchanged once every 3 years In order to keep the instrument in its original conditions please use exclusively distilled and filtered water 5 2 LED AND BUZZER INDICATIONS Green Red Buzze...

Page 25: ...rewed correctly Volume between H2 generator and consumer too big Pressure value could not be reached on time Increase value of timeout Verify capacity and consumption reduce consumption of GC 3 H2 Cel...

Reviews: