STANDARD USER MANUAL n° #0020-3 

SMU#0020-03 

 

Page 23/25 

INSTRUMENT INFORMATION: 

 

Instrument  information section provides  Firmware  Version, Instrument  Serial Number, Run time  and Cell 
total current. 

 

PRESSURE INFORMATION AND SETTINGS: 

 

Actual Pressure

 shows the pressure measured at the outlet of the instrument; 

Set Pressure

 is a parameter, where user could see and/or modify desired value of the pressure; 

Timeout

 is time in seconds showing for how long the 

Actual Pressure

 does not match the

 Set Pressure

Timeout Alarm

 is time in seconds given to the instrument after start to reach the 

Set Pressure

.  

 

IMPORTANT NOTE:

 For security reasons, the Instrument stops generation when the 

Timeout

 value became 

equal to 

Timeout Alarm

 

FLOW INFORMATION: 

 

 

Consumption

 shows current water consumption in ml/h; 

Production

 shows actual production of H

2

 in percentage from the maximum capacity;

 

Estimated Flow

  shows  actual H

2

  flow  in ml/min. Please  note  that this value  is not  directly measured but 

calculated inside the instrument. Use it only for estimation purposes. 

 

CELL INFORMATION: 

 

Cell information section provides Run time and Cell total current. 

Summary of Contents for HYDROXYCHROM

Page 1: ...e Application date Modified chapters 0 Creation 24 10 2012 All 1 Revision 25 10 2012 Ch 3 4 2 Revision 22 04 2015 Ch 4 6 2 3 Revision 14 12 2018 Ch 3 1 4 3 Editor S GUTIERREZ JP AMIET K TRAULLE Checke...

Page 2: ...IONING PRINCIPLE 8 CHAPTER 4 INSTALLATION AND OPERATION 9 4 1 RECEIPT OF INSTRUMENT AND CHECK 9 4 2 DELIVERY CONTENT 9 4 3 GENERATOR INSTALLATION 9 4 4 CONNECTIONS 10 4 4 1 Fluidic connections 10 4 4...

Page 3: ...the delivery date The replacement of the defective components will be free except the transport or travel fees that will be invoiced according the current tariffs CHROMATO SUD accepts no liability fo...

Page 4: ...operations alarms as well as troubleshooting It is applicable to the below generation models and software HYDROXYCHROM Viewer Version 1 0 produced since October 2012 HYDROXYCHROM 100 HYDROXYCHROM 160...

Page 5: ...cause 2 2 W E E E PRODUCT RECYCLING DECLARATION In agreement with the European EC 2002 96 directive on electrical and electronic equipment recycling this product may not be disposed in the garbage Fo...

Page 6: ...d water TOC free Conductivity 0 20 S cm Water capacity 5L tank outside 0 4L tank inside Water consumption 5L water generates about 6000L Hydrogen Safety Low H2 stored volume over pressure valve intern...

Page 7: ...ear panel left side has Mains plug with fuse and switch DB9 interface connectors for RS485 remote control USB dry relay The rear panel in the center has ZEROWATER input from external tank fitting for...

Page 8: ...YDROXYCHROM 100 and HYDROXYCHROM 160 are equipped with a single layer H2 cell H2 GENERATION MODULE O2 Outlet Water Inlet H2 Outlet External 5L Water Tank ZERO WATER box Water pump 1 Static dryer 24 H2...

Page 9: ...charged to the customer It is well recommended to keep the wood box for future shipments 4 2 DELIVERY CONTENT Quantity Description 1 Hydrogen Generator HYDROXYCHROM with its factory communication opt...

Page 10: ...at a maximum distance of 1 2 m from the generator It can nevertheless be placed at the same level or above the instrument The maximum difference in height between the generator and the ZEROWATER tank...

Page 11: ...it was not possible verify that the cable use has a sufficient section and has a ground wire 3X1 mm2 minimum Make sure laboratory differential circuit breaker can absorb an inrush current of at least...

Page 12: ...provides User Interface to control Gas Generator Network through USB interface 4 6 1 SPECIFICATIONS Requirements PC or Laptop under WinXP SP2 or higher Win Vista Win 7 At least 5Mb of free space on th...

Page 13: ...you must log the computer in administrator Remark If you want work with HydroxychromViewer in Russian start by installing the Cyrillic code page procedure below and then come back here only available...

Page 14: ...STANDARD USER MANUAL n 0020 3 SMU 0020 03 Page 14 25 Step 3 Choose destination folder or use default Step 4 Complete installation...

Page 15: ...chrom Viewer V1 0 when the installation is complete you have to do a commit cf procedure below If you just installed Hydroxychrom Viewer V2 0 when the installation is complete this window will appear...

Page 16: ...STANDARD USER MANUAL n 0020 3 SMU 0020 03 Page 16 25 CYRILLIC CODE PAGE INSTALATION Step 1 Run install_cyrillic_codepage exe Step 2 Select your preferred language for this installation...

Page 17: ...tallation When the installation is complete this window will appear clic on yes the computer will restart automatically saving the modification you just made Remark After the computer restarted you ca...

Page 18: ...EDURE FOR COMPUTER UNDER EMBEDDED WINDOWS Open the program named command prompt in the toollbar In this program write commitcf as shown in the picture below and press enter The program asks you to res...

Page 19: ...STANDARD USER MANUAL n 0020 3 SMU 0020 03 Page 19 25 CHANGING LANGUAGE Step 1 Open Application partition Step 2 Open Hydroxychrom folder Step 3 Open Hydroxychrom_v2 ini file with Notepad...

Page 20: ...ep 4 Modify the second line of the code LanguageFile Hydroxychrom_UK lng for English language LanguageFile Hydroxychrom_FR lng for French language LanguageFile Hydroxychrom_RU lng for Russian language...

Page 21: ...R MANUAL n 0020 3 SMU 0020 03 Page 21 25 And open it again USAGE Step 1 Connect USB Step 2 Start HYDROXYCHROM Viewer Step 3 Adjust the COM port number that is associated with USB port Step 4 Left Clic...

Page 22: ...f following Off No gas generation On Gas generation is running but will be stopped after power off Continuous On Gas generation is running but will be automatically started every time after power on R...

Page 23: ...ure does not match the Set Pressure Timeout Alarm is time in seconds given to the instrument after start to reach the Set Pressure IMPORTANT NOTE For security reasons the Instrument stops generation w...

Page 24: ...ter should be exchanged once every 3 years In order to keep the instrument in its original conditions please use exclusively distilled and filtered water 5 2 LED AND BUZZER INDICATIONS Green Red Buzze...

Page 25: ...rewed correctly Volume between H2 generator and consumer too big Pressure value could not be reached on time Increase value of timeout Verify capacity and consumption reduce consumption of GC 3 H2 Cel...

Reviews: