STANDARD USER MANUAL n° #0020-3
SMU#0020-03
Page 25/25
5.3.
A
LARMS
+
T
ROUBLESHOOTING
Alarm #
Alarm Name
Cause
Cure
1
Water Level
Internal tank could not be filled up to
maximum level detector.
Tube between external tank and
device is leaky.
External tank is empty, add
water.
Push tube in the INLET
WATER quick connector.
2
H
2
Output pressure
Actual H
2
pressure could not reach
set value, because there is a leak
inside your generator or on the line
between generator and GC.
H
2
flow consumed by your GC is
superior to the generator’s capacity.
Proceed to a leak check with
HYDROXYCHROM
Viewer
software and verify if fittings
are screwed correctly.
Volume
between
H
2
generator and consumer too
big. Pressure value could not
be reached on time. Increase
value of timeout.
Verify
capacity
and
consumption,
reduce
consumption of GC.
3
H
2
-Cell Voltage
Bad quality of water.
Deionisant bags inside internal and
external tank are not any more
efficient.
H
2
-cell is dry.
Replace
actual
water
exclusively by distilled and
filtered water.
Exchange deionisant bags.
Verify if water is correctly
circulating. Circulating pump
should not be stopped.
4
Internal
communication
One or several cable(s) inside device
is(are) disconnected or damaged
One board inside device is damaged.
If
a
cable
is
simply
disconnected
due
to
vibrations during shipment,
please switch of the device
and reconnect simply the
cable.
If one board is damaged,
please
call
your
local
distributor for repair.