STANDARD USER MANUAL n° #0020-3
SMU#0020-03
Page 24/25
CHAPTER 5.
MAINTENANCE, TROUBLESHOOTING
5.1.
R
EGULAR
M
AINTENANCE
The Hydrogen generator HYDROXYCHROM does not need regular maintenance; it needs ZEROWATER and
electrical power. If the instrument is used according to recommendations made by the constructor, the
deionizing bags (internal and external tanks) should be exchanged once every 1 year and the internal tank
filter should be exchanged once every 3 years. In order to keep the instrument in its original conditions,
please use exclusively distilled and filtered water.
5.2.
L
ED AND BUZZER INDICATIONS
Green
Red
Buzzer
ON
OFF
OFF
STBY (by user), REMOTE – STBY, ON, CONTINEOUS ON, REMOTE – ON
FAST
OFF
OFF
STBY with Warning
FAST
OFF
FLASH
ON with Warning, CONTINEOUS ON with Warning, REMOTE – ON with Warning
OFF
FAST
FAST
STBY with Error
FLASH
OFF
SLOW
LeakTest in process
FLASH
OFF
OFF
LeakTest success
FLASH
SLOW
FAST
LeakTest Fail
FAST
FAST
OFF
USB-CABLE-CONNECTED
Variable OFF
OFF
H2 Flow Production indication versus maximum capacity
STBY = STAND-BY