- 22 -
Minimum and maximum values which can be set for zero and full scale corrections
:
ANALOG OUTPUT TYPE
Minimum
Maximum
0÷10 V
-0.150
10.200
0÷5 V
-0.150
5.500
±
10 V
-10.300
10.200
±
5 V
-5.500
5.500
0÷20 mA
-0.200
22.000
4÷20 mA
-0.200
22.000
NOTE:
the analog output may also be used in the opposite manner, i.e. the weight setting that
corresponds to the analog zero (
) may be greater than the weight set for the analog full scale
(
). The analog output will increase towards full scale as the weight decreases; the analog
output will decrease as the weight increases.
For example:
= 10000
= 0 analog output 0÷10 V
Weight =
0 kg
analog output = 10 V
Weight = 5000 kg
analog output =
5 V
Weight =10000 kg analog output =
0 V
All analog outputs of the instrument are ACTIVE and SINGLE ENDED type, therefore
they can be connected only to PASSIVE receiver devices. The minimum load allowed for
voltage outputs is 10 kohm, the maximum load allowed for current outputs is 300 ohm.
SERIAL COMMUNICATION SETTING
-
/
: communication port.
-
: it disables any type of communication (default).
-
: MODBUS-RTU protocol; possible addresses: from 1 to 99 (see Communication
protocols manual).
-
: continuous weight transmission protocol to RIP5/20/60, RIP50SHA, RIPLED series
remote displays; the remote display shows the net weight or gross weight according to its
settings (set:
=
,
=
,
=
).
ACTIVE
ANALOG
OUTPUT
PASSIVE
RECEIVER
Voltage or current analog signal
-
COMMON
+