- 14 - 

PRESET TARE (SUBTRACTIVE TARE DEVICE) 

 

It is possible to manually set a preset tare value to be subtracted from the display 
value  provided  that  the 



  ≤  max  capacity  condition  is  verified.  In  multi-

interval instruments, the max permitted value is Max1 (max capacity of range 1). 

 
By default the instrument shows the last programmed preset tare value: to apply it press

 

 and 

then ENTER. 
After  setting  the  tare  value,  going  back  to  the  weight  display,  the  display  shows  the  net  weight 

(subtracting  the  preset  tare  value)  and  the  NET  symbol  lights  up  to  show  that  a  tare  has  been 
entered. To delete a preset tare and return to gross weight display, hold down  TARE for about 3 
seconds or keep the NET/GROSS input (if any) closed for the same length of time (3 seconds). The 
preset tare value is set to zero. The NET symbol is turned off when the gross weight is displayed 

once again. 

 

While the net weight is displayed, keep 

 

pressed to display the preset tare. When 

the key is released the net weight will be displayed again. 

 

-

 

 IF  A  SEMI-AUTOMATIC  TARE  (NET)  IS  ENTERED,  IT  IS  NOT  POSSIBLE  TO 

ACCESS THE ENTER PRESET TARE FUNCTION. 

-

 

IF  A  PRESET  TARE  IS  ENTERED,  IT’S  STILL  POSSIBLE  TO  ACCESS  THE  SEMI-

AUTOMATIC TARE (NET) FUNCTION. THE TWO DIFFERENT TYPES OF TARE ARE 

ADDED. 

 

ALL THE SEMI-AUTOMATIC TARE (NET) AND PRESET TARE FUNCTIONS WILL BE 

LOST WHEN THE INSTRUMENT IS TURNED OFF. 

 

 

SEMI-AUTOMATIC ZERO (WEIGHT ZERO-SETTING FOR SMALL VARIATIONS) 

 

By closing the SEMI-AUTOMATIC ZERO input, the weight is set to zero; alternatively, by pressing 
the 

 

key  for  less  than  3  seconds,  the 



  message  is  displayed  for  3  seconds,  by 

pressing ENTER the weight is set to zero. The zero-setting will be lost upon instrument power-off. 
This function is only allowed when: 

-

 

the weight is stable; otherwise the 



 alarm appears; 

-

 

the combined effect of the semi-automatic zero and zero tracking does not exceed 2% of the 
maximum capacity, otherwise the alarm 



 appears. 

 

 
 

 

0

 

+

 



 

Summary of Contents for TLM8

Page 1: ...ENGLISH ENGLISH ENGLISH ENGLISH User Manual version 1 04 TLM8 2014 30 UE EN55022 2010 EN61000 6 2 2005 EN61000 6 4 2007 SYSTEM IDENTIFICATION CE MAPPROVED EN45501 2015 2014 31 EU OIML R76 2006...

Page 2: ...er No guarantee against misuse Batteries Laumas provides 1 year guarantee from the date of delivery note against material defects or battery manufacturing faults Disposal of Waste Equipment by Users i...

Page 3: ...N IS AVAILABLE 11 ALIBI MEMORY READING 11 OUTPUTS AND INPUTS CONFIGURATION 12 SEMI AUTOMATIC TARE NET GROSS 13 PRESET TARE SUBTRACTIVE TARE DEVICE 14 SEMI AUTOMATIC ZERO WEIGHT ZERO SETTING FOR SMALL...

Page 4: ...DECLARATION OF CONFORMITY EU 31...

Page 5: ...ther cables connect it directly to the instrument terminal strip without breaking its route with support terminal strips Use RC filters on the instrument driven solenoid valve and remote control switc...

Page 6: ...relating to shocks and vibration windy conditions seismic conditions in the installation setting stability of the support structure EARTHING THE WEIGHED STRUCTURE By means of a copper wire with suita...

Page 7: ...the positive excitation wire and the negative excitation wire must be equal or similar to the one indicated in the load cell data sheet input resistance Load cell voltage measurement use a digital mu...

Page 8: ...ModBus R T U protocol remote display printer Optional integrated Profibus DP DeviceNet CANopen Profinet IO Ethernet IP Ethernet TCP IP Modbus TCP RS232 RS485 SERCOS III PowerLink CC Link output 8 inde...

Page 9: ...x 5 x 10 x 20 x 50 x 100 DIGITAL FILTER READINGS PER SECOND 11 levels 5 600 Hz RELAY OUTPUTS N 5 max 115 VAC 150 mA DIGITAL INPUTS N 3 optoisolated 5 24 VDC PNP SERIAL PORTS RS485 BAUD RATE 2400 4800...

Page 10: ...ysteresis Confirm or enter in submenu Setting general parameters press immediately followed by Setting preset tare press immediately followed by Load distribution display SYMBOL Function net weight se...

Page 11: ...7 MENU MAP Into menus changes are applied right after pressing the ENTER key no further confirmation is required SETPOINT SYSTEM PARAMETERS optional...

Page 12: ...8 3 s...

Page 13: ...fy the displayed figure or menu item to select a new figure or modify the displayed menu item ESC to cancel and return to the previous menu FILTER ON THE WEIGHT Setting this parameter allows a stable...

Page 14: ...esh frequency Hz 0 12 300 1 150 100 2 260 50 3 425 25 4 default 850 12 5 5 1700 12 5 6 2500 12 5 7 4000 10 8 6000 10 9 7000 5 A 6 600 The A filter can only be set if the instrument is connected to one...

Page 15: ...creases for stored records from 1 to 999999 and is automatically reset to 1 after reaching 999999 WARNING the alibi memory cannot store all the 999999 records the number of records that can be stored...

Page 16: ...ration mode is selected the following options are also active the contact will switch on the basis of gross weight the contact will switch on the basis of net weight If the net function is not active...

Page 17: ...wise the weight is displayed when the input is closed the data are sent for printing if in the communication protocol of either serial port the parameter is set If the alibi memory is active data stor...

Page 18: ...ed once again While the net weight is displayed keep pressed to display the preset tare When the key is released the net weight will be displayed again IF A SEMI AUTOMATIC TARE NET IS ENTERED IT IS NO...

Page 19: ...spond to the value set in the PLC program default calibration full scale E g if I am using a 4 20 mA output and in the PLC program I wish to have 20 mA 8000 kg I will set the parameter to 8000 analog...

Page 20: ...oad allowed for current outputs is 400 ohm SERIAL COMMUNICATION SETTING it disables any type of communication default MODBUS RTU protocol possible addresses from 1 to 99 see Communication protocols ma...

Page 21: ...minimum baud rate 2400 baud 40 Hz with minimum baud rate 4800 baud 80 Hz with minimum baud rate 9600 baud 100 Hz with minimum baud rate 19200 baud 200 Hz with minimum baud rate 38400 baud 300 Hz with...

Page 22: ...ct the above mentioned resistors DIRECT CONNECTION BETWEEN RS485 AND RS232 WITHOUT CONVERTER Since a two wire RS485 output may be used directly on the RS 232 input of a PC or remote display or printer...

Page 23: ...on on zero and the one previously stored weigh load percentage on each channel Load diagnostics the instrument with stable weight calculates and stores the load percentage on each channel If under nor...

Page 24: ...e corresponding output opens Setting ensure that the corresponding output closes Analog Output Option Test It allows the analog signal to range between the minimum and the maximum values starting from...

Page 25: ...distribution deletion equalization modification or deletion of the manual selection of active channels it deletes stored events confirmation is requested press ENTER to proceed or press ESC to cancel...

Page 26: ...icient alibi memory status SETPOINT PROGRAMMING From the weight display press MENU to access the setpoint setting MENU ENTER to enter a menu confirm the data entry to modify the displayed figure or me...

Page 27: ...the indicator CONNECTION TO THE WEIGHT INDICATOR Connection via RS485 port Weight indicator configuration Configure on the serial port used the protocol with the following parameters TLM8 multi chann...

Page 28: ...G the map shows only the additional menu items that are enabled on the W series instruments connected to the TLM8 Into menus changes are applied right after pressing the ENTER key no further confirmat...

Page 29: ...TLM8 display If the configuration is performed on the TLM8 the W series instrument connected must be restarted to allow synchronization REMOTE KEYPAD LOCKING keypad locked keypad unlocked ADDITIONAL...

Page 30: ...nd the permitted values press ESC to quit the setting mode leaving the previous value unchanged Examples a number of decimals is selected for full scale which exceeds the instrument s display potentia...

Page 31: ...BUS RTU ASCII __O F_ __O L_ __O F_ __O L_ __O F_ aa CR __O L_ RIP __O F_ __O L_ __O F_ __O L_ __O F_ __O F_ __O L_ HDRIP N _ERCEL _ER_OL _ER_AD _ER_OF O__SET CONTIN _ERCEL _ER_OL _ER_AD _ER_OF O__SET...

Page 32: ...BASE Addr 01 G 878 kg N 589 kg T 289 kg BASIC PRINTOUT ALIBI ENABLED TLM8 BASE Addr 01 ID 6 G 878 kg N 589 kg T 289 kg BASIC PRINTOUT PEAK ENABLED TLM8 BASE Addr 01 G 1204 kg N 831 kg T 373 kg P 2103...

Page 33: ...4 kg CH1 23 5 CH2 24 1 CH3 15 5 CH4 16 7 CH5 9 0 CH6 10 2 CH7 ERR load cell connected but in error CH8 OFF load cell not connected CURRENT AND STORED DISTRIBUTION TLM8 BASE Addr 01 STORED MEMRIP GROSS...

Page 34: ...phase has been completed by the manufacturer and includes all the tests listed by EN45501 2015 8 2 2 The second assessment phase must be completed by the authorities in charge in the Country in which...

Page 35: ...den Normen bereinstimmt F D claration de conformit Nous d clarons avec cela responsabilit que le produit auquel se rapporte la pr sente d claration est conforme aux normes cit es ci apr s CZ Prohl en...

Page 36: ......

Reviews: