- 22 -
-
: maximum capacity for single interval instruments
-
: maximum capacity of range 1 for multi-interval or multiple range instruments
-
: maximum capacity of range 2 for multi-interval or multiple range instruments
-
: maximum capacity of range 3 for multi-interval or multiple range instruments
-
: gravity acceleration value in the calibration area
-
: gravity acceleration value in the place of use
-
: weight correction depending on temperature
-
: restoring the printing/storage function
-
: valid unit of measure when the coefficient function is active
-
: weight display change coefficient
-
: alibi memory status
SETPOINT PROGRAMMING
From the weight display, press MENU to access the setpoint setting.
MENU/ENTER: to enter a menu/confirm the data entry.
: to modify the displayed figure or menu item.
: to select a new figure or modify the displayed menu item.
ESC: to cancel and return to the previous menu.
-
(from 0 to max full scale; default: 0): Setpoint; relay switching occurs when the weight
exceed the value set in this parameter. The type of switching is settable (see section
OUTPUTS
AND INPUTS CONFIGURATION
).
-
(from 0 to max full scale; default: 0): Hysteresis, value to be subtracted from the setpoint
to obtain contact switching for decreasing weight. For example with a setpoint at 100 and
hysteresis at 10, the switching occurs at 90 for decreasing weight.
…
…