- 11 -
ALIBI MEMORY (ONLY FOR INSTRUMENTS WHERE THIS OPTION IS AVAILABLE)
To save a weight to the alibi memory press the key
for less than 3 seconds or close the
dedicated input or send the command via serial protocol (see the Protocols manual).
If the net function is active, the net weight is stored; otherwise, the gross weight is stored.
Saving to the alibi memory is only possible if:
-
the alibi memory is enabled;
-
the weight is stable;
-
the weight is above the minimum weight (20 e);
-
the weight is below the max capacity;
-
the weight has changed since the latest storage to alibi.
An identifying number is automatically assigned to each record stored in the memory, it increases
for stored records from 1 to 999999, and is automatically reset to 1 after reaching 999999.
WARNING: the alibi memory cannot store all the 999999 records; the number of records that
can be stored depends on the instrument type and on the program used.
The alibi memory is used in a “circular” manner: when the memory end is reached, the system will
restart from the memory beginning by overwriting the first record; an alarm message is displayed
(
).
ALIBI MEMORY READING
(identification): enter the ID number of your required record and confirm, the display will show:
-
Gross weight (
) or net weight (
); confirm with ENTER;
-
Tara weight (
) or preset tare (
), only if the recalled weight is a net weight; confirm
with ENTER;
The parameter provides the ID of the last record stored in Alibi Memory by default.
If the entered identifier does not match any record, the message
is displayed.
+