Proline Low-temperature Thermostats
42
Starting up
YACE0072 / 21.08.07
Continuation of submenu
LMDkt
Æ
O0c--
(Master): Setting external control
parameters
* Important:
The windows for setting the
additional parameters for the
external control for the
Master
Controller:
Ds
(=tve) and
Dc
(=tde),
and for the
Slave Controller
:
Da
(=P-controlller),
are only displayed when
D
has
been changed from the works
setting of "Automatic" to
"Manual".
Changeover should only be
carried out by experienced
control technicians.
End
Dlc
°C
The works setting (default)
of the external control
parameters has been
restored and
cmlD
is
displayed.
longer
than
3 sec.
Control works settings
cDE
°C
Display flashes. Set
with or .
Range: 0.0K … 5.0K.
Corr.variable limits [K]
Dr/K
°C
longer
than
3 sec.
Display flashes. Changeover
tne, tve and tde from
@
=Automatic to
LM
=Manual
with or .
Auto/man. for tne, tve, tde
D @
°C
Return to the
previous page