Proline Low-temperature Thermostats
YACE0072 / 21.08.07
Interfaces - Modules
109
8.4 Analogue
module
The analogue module (order no. LRZ 912) has 2 inputs and 2 outputs, which are brought out on a 6-
pole DIN socket to Namur Recommendation (NE28). The inputs and outputs can be set independently
as 4…20 mA, o…20 mA or 0…10V interface. Various functions can be selected for the inputs and
outputs. Accordingly, the signal on the input is interpreted differently and different information is output
via the output connection.
In addition the interfaces can be scaled freely according to the set function.
For measuring transducer are 24 V DC available.
The following values can be specified via the inputs:
-
Setpoint temperature with function:
LM rR
or
Set temperature
.
-
External actual temperature with function:
LM rD
or
ext. actual temperature
.
-
Pump power with function:
LM OO
or
Pump power
.
The following values can be specified via the outputs:
-
Setpoint temperature with function: Master:
LM rR
or Command:
Set temperature
.
-
The temperature source with which active control occurs:
LM rB
Controlled temp.
.
-
Actual temperature (bath temperature):
LM r0
or
Internal Temp.
.
-
External actual temperature from Pt100:
LMrDO
or
Temp.external Pt100
.
-
External actual temperature from analogue input:
LMrD@
or
Temp.external analogue
.
-
External actual temperature from the serial interface:
LMrDR
or
Temp.external serial
.
-
Actuating
signal:
LM X
or
Actuating signal
.
-
Pump
power:
LM OO
or
Pump power
.
-
Pump
speed:
LMrDl
or
Pump speed
.
In addition the interfaces can be scaled freely with
K </
/
G 0/</
in % or
minimal value
/
maximal value
according to the set function.
For example:
4 mA corresponds to 0 °C and 20 mA corresponds to 100 °C.
−
Accuracy of the inputs and outputs after calibration better than 0.1% F.S.
−
Inputs,
current
−
Inputs, voltage
−
Outputs,
current
−
Outputs,
voltage
Input resistance < 100 Ohm
Input resistance > 50 kOhm
Burden < 400 Ohm
Load > 10 kOhm
Connection of the analogue inputs and outputs
A 6-pole round connector with screw locking and contact arrangement according to DIN EN 60130-9
or IEC 130-9 is needed.
A suitable coupling plug can be obtained under order no. EQS 057.
View of the socket (front) or solder side of plug:
Socket 71S (till end 2006)
Socket 74S (from 2007 on)
Pin 1
Output 1
Pin 2
Output 2
Pin 3
0 V reference potential
Pin 4
Input 1
Pin 5
0 V reference potential
Pin 6
Input 2
Pin 1
Output 1
Pin 2
Output 2
Pin 3
0 V reference potential
Pin 4
Input 1
Pin 5
+24 V (max. 0,1 A)
Pin 6
Input 2
Use shielded lines. Connect shielding with connector housing!
Summary of Contents for RP 1290
Page 2: ......