Proline Low-temperature Thermostats
YACE0072 / 21.08.07
Starting up
81
−
If the field in the column "Pump" has a
black background, the entry mode for the
Pump level
is obtained by pressing the
key
.
−
With
or
select Pump Level 1 – 8
or ------- and confirm with
.
------ stands for "no change to previous
segment", i.e. when “------“ is present in
all fields, the pump level always retains
the start setting or the setting before the
program start.
−
Continue
with to the field "Out 1",
"Out 2" or "Out 3".
−
The contact outputs of the contact
module (if present, special accessory)
are programmed here.
−
If the field in the column "Out 1" has a
black background, the entry mode for the
Contact output
is obtained by pressing
the
key.
−
With
or
select
-------
,
Open
or
Closed
and confirm with
.
-------
stands for no change with respect
to the previous segment, i.e. if
-------
is
present in all fields, the contact setting of
the start setting or that from the program
start is retained.
−
If applicable, continue with
to "Out 2"
and "Out 3".
−
Programming is terminated with
or
End .
Help Menu
End T
set
T
fix
Level 8
Level 7
Level 6
Level 5
Level 4
Level 3
Level 2
Level 1
-------
Pump level
Help Menu
End T
set
T
fix
Contact out
-------
open
closed