Technical data
13.04.99
YACE0059
25
9
Technical data (to DIN 58966)
Low-temperature thermostats
RE 104
RE 106
RE 107
RE 112
RE 120
Operating temperature range
°C
-10 to 120
-20 to 120
-35 to 120
-30 to 120
-30 to 120
Ambient temperature range
°C
5...40
(
#
Section 5.1)
Setting resolution
°C
0.1
Indication resolution
°C
0,1
Indication accuracy
±0.4°C ±0.5% additive calibration from indicated value
➃
Temperature control
± °C
0,05
Safety features
➀
FL
Heater
power
230 V; 50/ 60 Hz
115 V; 60 Hz
100 V; 50/60 Hz
kW
1,5
1.3
1.0
Cooling capacity
(eff.) @ with ethanol
at 20 °C ambient
temperature
20°C
0°C
-20°C
-30°C
kW
0.15
0.10
0.05*
----
0.20
0.15
0.05
----
0.30
0.22
0.10
0.06
0.30
0.23
0.13
0.04
0.35
0.25
0.10
0.04
Pump type
pressure pump with choice of 5 output steps
Max. discharge pressure
➁
bar
0.4
Max. flow rate
➁
l/min
17
Pump connections
mm
nipples 13 mm dia.
Max. bath volume
l
3 to 4.5
4 to 6
4 to 6
9 to 12
14 to 20
Bath opening (W x D)
mm
130x105
150x130
150x130
200x200
300x350
Bath depth
mm
160
160
160
200
160
Usable depth
mm
140
140
140
180
140
Height top edge of bath
mm
363
396
396
441
441
Overall size (W x D x H)
mm
180x320x524
200x400x557
200x400x557
250x400x602
350x530x602
Weight
kg
21
24
26
30
34
Power
con-
sumption
230 V; 50/ 60Hz
115 V; 60 Hz
100 V; 50/60 Hz
kW
1.7
1.4
1.1
1.8
1.4
1.1
1.9
1.4
1.1
1.9
1.4
1.1
2.0
1.4
1.2
Ref. No.
230 V; 50/60 Hz
LCK 0864
230 V; 50 Hz
LCK 1861
LCK 1867
LCK 1870
LCK 1873
115 V; 60 Hz
LCK 4861
LCK 4864
LCK 4867
LCK 4870
LCK 4873
100 V; 50/60 Hz
LCK 6861
LCK 6864
LCK 6867
LCK 6870
LCK 6873
*@ -10°C
➀
FL: suitable for flammable and non-flammable liquids
➂
➩
Section 7.2 last item
➁
at pump output step 5
Units to EU Directive 89/336/EWG ( EMC ) and 73/ 23/ EWG (low-voltage) with CE-mark
We reserve the right to make technical alterations!