Instruction Manual

HI935005 - HI935005N

HI935002 - HI935009

 

Portable 

K-Thermocouple 

Thermometers 

www.hannainst.com

Summary of Contents for HI 935002

Page 1: ...Instruction Manual HI935005 HI935005N HI935002 HI935009 Portable K Thermocouple Thermometers www hannainst com ...

Page 2: ...which have been designed using the latest microprocessor technology to provide reliable and accurate measurements in a wide temperature range Standard features include K type probes waterproof casing dual level LCD low battery detection long battery life and two year warranty HI935005 has been designed for laboratory and field use while HI935005N is suitable for high accuracy meter and probe can b...

Page 3: ...ear the HI and LO values BL HI935005N only toggle the backlight ON and OFF CAL HI935005N only press and hold for 5 seconds to enter the Calibration mode with reading within 3 C range press to abort the Calibration mode Range may be limited by probe SPECIFICATIONS Range 50 0 to 199 9 C 200 to 1350 C 58 0 to 399 9 F 400 to 2462 F Resolution 0 1 C up to 199 9 C 1 C outside 0 1 F up to 399 9 F 1 F out...

Page 4: ...ement DISP select Normal Relative or T1 T2 measuring mode ALT HI935009 only enable the second function keys the ALT tag turns on to indicate that the second functions are enabled Note The ALT key can be released before pressing the second function key for 1 hand operation ALT CAL press and hold for about 5 seconds to enter the Calibration mode with reading within 3 C range ALT AOFF set auto power ...

Page 5: ...d updating Hi Lo and relative values In HI935009 only the held value is also stored in non volatile memory and can be recalled by pressing ALT MR key While the MR key is held the HOLD tag lights up and the meter displays the stored value when the MR key is released after 1 second the meter returns to normal measurement mode HIGH LOW TEMPERATURES The maximum and minimum temperatures are continuousl...

Page 6: ...e corresponding tags will light up to inform the user To select the desired channel use the T1 T2 key The 2 channel models are also provided with the DISP function which allows the user to select the information to be displayed While in T1 or T2 pressing DISP will switch the display between standard and relative measurement modes In the standard mode the main body of the LCD shows the current temp...

Page 7: ...ce made from distilled water Immerse the temperature probe in the center of the ice bath taking care not to touch the ice with the probe tip Ensure that the meter is measuring a temperature within 3 C To enter the Calibration mode press and hold the CAL button for about 5 seconds The CAL tag turns on to indicate that the Calibration mode has been entered Note If the measurement is outside the 3 C ...

Page 8: ...unction with the HI766HD probe handle HI766PA Roller surface probe max 320 C 600 F HI766PB Surface probe max 650 C 1200 F HI766PC Penetration probe max 900 C 1650 F HI766PD Air probe max 300 C 570 F HI766PE1 General purpose probe max 900 C 1650 F HI766PE2 General purpose probe max 900 C 1650 F grill surface probe HI766B4 Grill surface probe with 70 cm 27 6 cable protect ed with stainless steel jac...

Page 9: ...truments Inc Highland Industrial Park 584 Park East Drive Woonsocket RI 02895 USA www hannainst com Local Office Hanna Instruments Inc Highland Industrial Park 584 Park East Drive Woonsocket RI 02895 USA Phone 800 426 6287 Fax 401 765 7575 e mail tech hannainst com ...