Maintenance
13.04.99
YACE0059
23
8 Maintenance
8.1 Cleaning
Before cleaning the unit, pull out the mains plug!
The unit can be cleaned with water adding a few drops of detergent (washing up liquid), using a moist
cloth
Water must not enter the control unit
☞
−
Carry out appropriate detoxification if dangerous material has been spilled on or
inside the unit.
−
Method of cleaning and detoxification are decided by the special knowledge of the
user. In case of doubt please contact the manufacturer
8.2
Maintenance and repair
−
Before any maintenance and repair work pull out the mains plug!
−
Repairs on the control unit must only be carried out by properly qualified personnel!
LAUDA thermostats are largely maintenance-free. If the thermostating liquid becomes dirty it has to be
replaced (
➾
Section 5.2.).
☞
−
If a fuse blows (
!
supply indication not alight) fit only fuses as specified (2 x T 16 A; 1 x
T 2,5 A, size 5 x 20
!
fuses are inside the unit.
F3; T 2,5 A
UL 487-1
F1; T 16 A
F2; T 16 A