In each case, the heat transfer liquids cover a specific tem-
perature range. This temperature range must match the tem-
perature range of your application.
The use of heat transfer liquids can cause hazards from high or
low temperatures and fire if certain temperature thresholds are
exceeded or undercut or if the container breaks and there is a
reaction with the heat transfer liquid .
The heat transfer liquid safety data sheet specifies all possible
hazards and appropriate safety measures for handling the
liquid. The safety data sheet must therefore be consulted for
the intended use of the device.
1.6
Materials
All parts coming into contact with the heat transfer liquid are made
of high quality materials suitable for the operating temperature.
Stainless steel and temperature-resistant plastics are used.
1.7
Hoses
Only LAUDA hoses are permitted to be used for the external
hydraulics circuit. LAUDA hoses are hoses that are approved by
LAUDA DR. R. WOBSER GMBH & CO. KG. When selecting suit-
able hoses for the application, the permissible temperature range
and the maximum permissible pressure must be particularly
observed.
1.8
Application area
The device is exclusively permitted to be used in the following
areas.
Commercial area
Indoor use, no outdoor installation
Ambient temperatures from 5 to 40 °C
1.9
Personnel qualification
Operating personnel
Operating personnel are employees that have been instructed by
technical staff in the intended use of the device according to the
operating manual.
Safety
Microcool
8
Summary of Contents for MC 250
Page 39: ...Fig 12 Menu Operation Microcool 39...
Page 56: ...Indication on the display Description NTC sensor break NTC short circuit Faults Microcool 56...
Page 67: ......
Page 68: ......
Page 70: ......