4.
Drain the condenser afterwards.
Further information about draining the condenser
can be found in .
5.
Reconnect the device to the water supply and rinse it thor-
oughly.
Allow at least 10 litres of water to flow through the
device.
7.7
Checking the heat transfer liquid
Soiled or degenerated heat transfer liquid must be replaced. Fur-
ther use of the heat transfer liquid is only permitted with appro-
priate test results.
The heat transfer liquid must be checked according to DIN 51529.
Maintenance
Microcool
53