Regardless of the current settings, all data is saved with European
date spelling, in the 24-hour format.
In the Graph settings menu, you can modify the time interval
between individual protocol inputs as required. With the
[Recording
period]
option, you can also set one of the pre-defined time inter-
vals.
7.8 Viewing the device status
1.
Open the home window.
2.
Select the
Settings
Device status
tiles.
The
Device status
menu shows product line and device type of
the constant temperature equipment. You can also call up the
following data:
For the error analysis, the constant temperature equipment stores the last
warning, error and any alarm messages that occurred. The content of this
error memory is displayed as a list of messages. The number of messages that
can be stored depends on the device.
1.
Tap the
Error memory
tile.
You will see the list of messages; the most recent message is at
the top.
2.
Push the scrollbar button to scroll down the list to the required input.
You will see the following information for each message:
No.
: Shows a consecutive message number which is given to
each message chronologically.
Source
: Shows the name of the module that triggered the mes-
sage.
Code
: Displays the encoded description of the alarm, warning or
error.
Date
/
time
: Shows the exact time of the message.
Subcode
: Shows additional information about the message.
Device data cannot be modified and is only used to support the LAUDA
service when carrying out a diagnosis.
1.
Tap the
Device data
tile.
Among other things, temperature values of the constant tem-
perature equipment and the application are displayed, as well as,
where applicable, characteristic values and actuating variables of
the pump, heating and cooling functions.
Read the error memory
Display device data
V1
Command Touch operating unit
42 / 51