5.11.3
Write commands
Write commands are data specifications from the PC to the ther-
mostat.
Command
Description
OUT_SP_00_XXX.XX
Desired value transfer with maximum 3 places in front of the decimal
point and a maximum of 2 places after.
OUT_SP_04_XXX
Write [Hi] Outflow temperature limit upper value
OUT_SP_05_XXX
Write [Lo] Outflow temperature limit lower value
OUT_PAR_00_XX.X
Configuration of the control parameter Xp
OUT_PAR_01_XXX
Configuration of the control parameter Tn (5
-
180 s; 181 = Off)
OUT_PAR_02_XXX
Configuration of the control parameter Tv
OUT_PAR_03_XX.X
Configuration of the control parameter Td
OUT_MODE_00_X
Keyboard master: 0 = free / 1 = blocked (corresponds to: "KEY")
START
Turns the device on (from standby).
STOP
Switches the device to standby (switches the pump and Peltier ele-
ments off).
Take the following information into account:
" " (blank space) can also be used for "_".
Response from thermostat will be "OK" or, if there is an error,
"ERR_X".
Acceptable data formats
-XXX.XX
-XXX.X
-XXX.
-XXX
XXX.XX
XXX.X
XXX.
XXX
-XX.XX
-XX.X
-XX.
-XX
XX.XX
XX.X
XX.
XX
-X.XX
-X.X
-X.
-X
X.XX
X.X
X.
X
-.XX
-.X
.XX
.X
5.11.4
Read commands
Read commands are data demands from the PC to the thermostat.
Command
Description
IN_PV_00
Query outflow temperature
IN_SP_00
Query set temperature
IN_SP_04
Query of outflow temperature limit Hi
Operation
LOOP
33
Summary of Contents for L 100