7
Technical data
The information has been conveyed in accordance with
DIN 12876.
Specification
Unit
Loop 100
Loop 250
Working temperature range
°C
4 – 80
Temperature stability
K
±0.1
Ambient temperature range
°C
5 – 40
Storage temperature
°C
5 – 40
Power supply
---
100 – 240 VAC; 50/60 Hz
Power consumption
W
240
400
Heating output at ambient
temperature 20 °C
‑
T
Water
= 20 °C
W
> 100
> 250
Cooling capacity at ambient
temperature 20 °C
‑
T
Water
= 20 °C
W
120*
250*
‑
T
Water
= 10 °C
W
60*
130*
Pump type
---
Delivery pump
‑
Discharge pressure
bar
0.8
‑
Flow rate
L/min
2.6
‑
Tank content
ml
300
Hose connections
---
Quick connections ¼ ”
Hose nozzle/CPC in-line socket type PMC 2204 (EOA 077)
Display
---
OLED
‑
Resolution
Pixels
128 x 64
‑
Size
mm
61.4 x 30.7
‑
Setting resolution
°C
0.1
‑
Resolution of indication
°C
0.1
Data entry
---
Using three buttons (soft buttons)
Electric interface
---
RS 232 interface
Class division according to
DIN 12 876-1 for labor devices
‑
Class designation
---
I
‑
Identification code
---
NFL (non-flammable liquids)
Protection level (IP code)
according to IEC 60529
---
IP 21