Dual Scale Digital Clinical Thermometer

PLEASE READ CAREFULLY BEFORE USING

BODY TEMPERATURE

Normal body temperature can vary from 97

o

F (36.1 

o

C) to 99 

o

F (37.2 

o

C) . The generally accepted

“Normal” temperature is 98.6 

o

F (37.0 

o

C). Usually your temperature is lower in the morning than in

the afternoon. Normal temperature varies according to the location of the temperature reading. While
the oral temperature follows the guidelines previously described, a rectal temperature is generally 1.0

o

F (0.5 

o

C) higher. Conversely, an axillary (under the arm) temperature will be 1.0 

o

F (0.5 

o

C) lower.

GENERAL CARE AND USE

BEFORE USE PLEASE NOTE

1. The thermometer should be held during use. The patient should be seated or lying down.

2. TEMPERATURES SHOULD BE TAKEN UNDER ADULT SUPERVISION.

SPECIAL PRECAUTIONS

3. Avoid exposure to temperature extremes, dust, excessive humidity and direct sunlight.

4. Do not soak the thermometer in water or other liquids. Keep parts other than the probe away

from moisture.

5. Do not attempt to sterilize with heat or by boiling.

6. Do not bend or drop the unit. It is not shock proof.

7. Do not allow the unit to come into contact with any thinners or chemical solvents.

8. Do not attempt to disassemble the unit except to replace the battery.

9. Do not bite the main body or sensor unit.

DIRECTIONS  FOR  CLEANING

Clean the Thermometer probe before and after use by wiping it with soap and water. Rinse with 70%
Isoprophyl Alcohol,  then  rinse  thoroughly  in  cold  water.

SPECIFICATIONS

DISPLAY  RESOLUTION

1/10  DEGREE FOR BOTH 

O

C & 

O

F

MEASUREMENT RANGE

89.6 

O

F to 109.4 

O

F (32.0 

O

C to 43.0 

O

C)

ACCURACY

±0.3 

O

F / ±0.2 

O

C (89.6 

O

F to < 95.0 

O

F / 32 

O

C to < 35 

O

C)

±0.2 

O

F / ±0.1 

O

C (95.0 

O

F to < 104.0 

O

F / 35 

O

C to 40 

O

C)

±0.3 

O

F / ±0.2 

O

C (>104.0 

O

F to 109.4 

O

F / >40 

O

C to 43 

O

C)

MEASUREMENT METHOD

PEAK HOLD SYSTEM

BATTERY

ONE 1.5V DC ALKALINE MANGANESE BUTTON SIZE

LR41 TYPE OR EQUIVALENT

BATTERY LIFE

UPTO 3 YEARS UNDER NORMAL USE

AUTOMATIC SHUT-OFF

AFTER APPROXIMATELY 5 MINUTES

OPERATING TEMPERATURE

16 

O

C to 40 

O

C (61

O

F to 104 

O

F)

RELATIVE HUMIDITY

15% to 95% NON CONDENSING

STORAGE TEMPERATURE

-20 

O

C to 50 

O

C (– 4 

O

F to 122 

O

F)

RELATIVE HUMIDITY

15% to 95% NON CONDENSING

CAUTION

NEVER

 SUBMERGE THIS THERMOMETER IN ANY LIQUID.

NEVER

 ATTEMPT TO STERILIZE WITH HEAT, BY BOILING OR IN A GAS OR STEAM

AUTOCLAVE.

Thermometer performance may degrade in the following cases :

(a) Operation/storage  outside  the  specified  operating/storage  temperature

range  respectively.

(b) Mechanical  Shock/dropping.
(c) Patient  temperature  is  below  ambient  temperature.

OPERATING  INSTRUCTIONS

1. Ensure that the thermometer has been turned off before taking temperature.

2. The unit is factory set to 

o

F. To start the unit in 

o

F, press the blue coloured

ON/OFF switch momentorily once. To start the unit in 

o

C, press and hold the

blue coloured ON/OFF switch for approximately 2 Sec. If you need to go
back of 

 o

F scale (from 

o

C) press and hold the blue coloured ON/OFF switch

for approximately 2 seconds.

3. The following appear in quick sequence.

(a) This is the LCD display test (See Fig. 1). It appears for approximately 2

seconds  and  indicates  all  the  segments  of  the  LCD  are  functioning
properly.

(b) If the previous reading is between 89.6 

o

F ~ 109.4 

o

F, (32

o

C ~ 43

o

C), the

unit will display the last time measured temperature value for about 1
seconds with symbol M (Fig.2.) or the unit display Lo 

o

F with M sym-

bol (Fig. 3) for about 1 second after a battery replacement. Followed it
will displays 97.7 

o

F for about 1 second without symbol M (Fig. 4)

(c) The unit goes through the internal self checking test procedure. If the

unit is bad it will display “Err” (Fig. 5).

Note :

 When the temperature is below 50 

o

F (10 

o

C) or above 90 

o

C

(122 

o

F), it will display “Err”. It will auto power off in 10 seconds.

(d) When “Lo” appears with the flashing 

o

F (

o

C) sign (Fig. 6), the unit is

ready to take temperature.

Digital Display Window

Battery Cap

Sensor

On / Off Switch

E

M

F

M

Fig. 1

Fig. 2

Fig. 3

Fig. 4

Fig. 5

Fig. 6

F

F

F

ORAL  USE

1. Close mouth for about 5 minutes before measurement.

2. Place the thermometer tip under the tongue on a heat pocket (

3

) as shown in

Fig. 7.

3. Hold the thermometer in the same spot for about 40 seconds.

4. Readings will continue to increase and the 

o

F (or 

o

C) symbol will flash dur-

ing measurement (See Fig.8).

5. When peak temperature is reached, the 

o

F (or 

o

C) symbol will stop flashing

(See Fig. 9) and

a)

When the temperature is higher than 99.4 

o

F (or 37.4 

o

C) the beeper will

sound “beep-beep-beep” 8 times to indicate a high temperature.

b)

When the temperature is lower than or equal to 99.4 

o

F (or 37.4 

o

C) the

beeper will sound “beep” 4 times.

6. After measurement, record the temperature, then turn off the unit to save the

battery life. If you forget to turn it off, the unit will power down automatically
after approximately 10 minutes.

NOTE

 : THE MOUTH MUST REMAIN CLOSED DURING THE MEASUREMENT TO ENSURE

ACCURATE READINGS.

RECTAL USE

Rectal temperature should only be taken when oral and axillary temperatures are inconvenient. It is
recommended that axillary temperatures be taken on infants. LUBRICATE WITH LUBRICATING
JELLY AND INSERT THE PROBE NO MORE THAN 1/2 INCH INTO RECTUM. Hold thermometer
in place until you hear the beep signal and the 

o

F (or 

o

C) sign stops flashing (about 40 seconds.)

AXILLARY USE

Dry armpit before measurement. Place the tip of the probe under arm so that the tip is touching the skin.
This area is called axilla. Hold the thermometer so that it is parallel to the patient’s body and fold the
arm over the chest. This ensures that the tip of the probe is well covered and not influenced by air.
Hold thermometer in place until you hear the beep signal and the 

o

F (or 

o

C) sign stops flashing (about

3 minutes)

BATTERY REPLACEMENT

Replace battery with an alkaline LR 41 or equivalent type whenever the low battery indicator 

+ -

flashing.

1. Remove screw located on the back of the thermometer.

Press down on the battery cover and push in the direc-
tion  show  by  the  arrow  in  the  illustration.

2. Remove battery with a non-metallic probe.

3. Insert new battery. NOTE : The positive (+) side must face

outward.

4. Replace the cover.

5. Dispose off used battery properly.

NOTE : 

 Inserting battery with reverse polarity may damage

the unit. Dispose off this thermometer properly as per local
regulations.  This  Digital  clinical  thermometer  is  pre-
calibrated and can not be re-calibrated.

DISPOSABLE  PROBE  COVER  USE.

1. Insert the thermometer into probe cover. Take care not to push the

thermometer tip through the end of the probe cover.

2. Holding the thermometer and tab of the probe cover as shown, pull

away the covering from longer side.

3. Tilt thermometer to remove cover from shorter side.

4. Dispose off the probe cover after taking temperature.

NOTE : For rectal use, after applying probe cover lubricate thermometer

tip with a water soluble jelly.

LIFETIME GUARANTEE

Your  new  Faichney  Dual  Scale  Digital  Clinical  Thermometer  is  guaranteed  to  be  free  from
manufacturer’s defects for life under normal use.

However, if for any reason (other than abuse or normal battery replacement) your thermometer does
not function properly we will either repair or replace it free of charge for life.  This warranty is valid
only if unit returned for repair is properly packed to avoid damage in shipping.  All parts are covered
by the warranty, except batteries and probe covers, where applicable.

Please check the battery before returning it. This is usually the source of the problem, should repair
be needed, ship freight paid to :

Conforms to ASTM   E 1112

Fig.7

1.

2.

3.

7122301  Rev-01

Faichney  Medical  Company

17517,  Fabrica  Way  Suite  G
Cerritos  CA  90703, USA

Faichney  Neonatal  Thermometer

Reorder  No.  6202