OPERATING  INSTRUCTIONS

1. Ensure that the thermometer has been turned off before taking temperature.

2. The unit is factory set to 

o

F. To start the unit in 

o

F, press the blue coloured

ON/OFF switch momentorily once. To start the unit in 

o

C, press and hold

the blue coloured ON/OFF switch for approximately 2 Sec. If you need to
go back of 

 o

F scale (from 

o

C) press and hold the blue coloured ON/OFF

switch for approximately 2 seconds.

3. The following appear in quick sequence.

(a) This is the LCD display test (See Fig. 1). It appears for approximately

2 seconds and indicates all the segments of the LCD are functioning
properly.

(b) If the previous reading is between 89.6 

o

F ~ 109.4 

o

F, (32

o

C ~ 43

o

C),

the  unit will display the last time measured temperature value for
about  1  seconds  with  symbol  M  (Fig.2.)  or  the  unit  display  Lo 

o

F

with M symbol (Fig. 3) for about 1 second after a battery replacement.
Followed it will displays 97.7 

o

F for about 1 second without symbol

M (Fig. 4)

(c) The unit goes through the internal self checking test procedure. If the

unit is bad it will display “Err” (Fig. 5).

Note :

 When the temperature is below 50 

o

F (10 

o

C) or above 90 

o

C

(122 

o

F), it will display “Err”. It will auto power off in 10 seconds.

(d) When “Lo” appears with the flashing 

o

F (

o

C) sign (Fig. 6), the unit is

ready to take temperature.

ORAL  USE

1. Close mouth for about 5 minutes before measurement.

2. Place the thermometer tip under the tongue on a heat pocket (

3

) as shown

in Fig. 7.

3. Hold the thermometer in the same spot for about 40 seconds.

4. Readings  will  continue  to  increase  and  the 

o

F  (or 

o

C)  symbol  will  flash

during measurement (See Fig.8).

Fig.7

5. When peak temperature is reached, the 

o

F (or 

o

C) symbol will stop flash-

ing (See Fig. 9) and

a)

When the temperature is higher than 99.4 

o

F (or 37.4 

o

C) the beeper

will sound “beep-beep-beep” 8 times to indicate a high tempera-
ture.

b)

When the temperature is lower than or equal to 99.4 

o

F (or 37.4 

o

C)

the beeper will sound “beep” 4 times.

6. After measurement, record the temperature, then turn off the unit to save

the  battery  life.  If  you  forget  to  turn  it  off,  the  unit  will  power  down
automatically after approximately 10 minutes.

NOTE

 : THE MOUTH MUST REMAIN CLOSED DURING THE MEASUREMENT TO EN-

SURE ACCURATE READINGS.

RECTAL USE

Rectal temperature should only be taken when oral and axillary temperatures are inconvenient.
It  is  recommended  that  axillary  temperatures  be  taken  on  infants.  LUBRICATE  WITH
LUBRICATING  JELLY AND  INSERT THE  PROBE  NO  MORE  THAN  1/2  INCH  INTO
RECTUM. Hold thermometer in place until you hear the beep signal and the 

o

F (or 

o

C) sign stops

flashing  (about  40  seconds.)

AXILLARY USE

Dry armpit before measurement. Place the tip of the probe under arm so that the tip is touching the
skin. This area is called axilla. Hold the thermometer so that it is parallel to the patient’s body
and fold the arm over the chest. This ensures that the tip of the probe is well covered and not
influenced by air. Hold thermometer in place until you hear the beep signal and the 

o

F (or 

o

C) sign

stops flashing (about 3  minutes)

*********************************************************************

7022301 - Rev.01