Unit
IN 2560 XTW
IN 2560 XTW
Pump level
-40 °C
kW
8.70
9.50
4
-50 °C
kW
5.00
5.40
4
-60 °C
kW
3.00
3.30
4
Cooling water flow rate
l/min
29.0
29.0
---
Cooling water temperature
minimum/maximum
°C
10/30
10/30
---
Recommended cooling water
temperature
°C
15
15
---
Recommended cooling water
pressure difference
bar
3
3
---
Minimum cooling water pres-
sure difference*
bar
0.8
0.8
---
Maximum cooling water pres-
sure difference
bar
10
10
---
Maximum cooling water pres-
sure
bar
10
10
---
* only approved for a max. cooling water temperature of 20 °C
Table 54: Cooling output
Unit
IN 2050 PW
Pump level
IN 2560 PW
IN 2560 PW
Pump level
Cooling output
(ambient tem-
perature 20 °C)
at 50/60 Hz
at 50 Hz
at 60 Hz
140 °C
kW
20.00
8
25.00
28.00
8
100 °C
kW
20.00
8
25.00
28.00
8
20 °C
kW
20.00
8
25.00
28.00
8
10 °C
kW
15.00
8
25.00
28.00
8
0 °C
kW
10.80
8
24.50
27.50
8
-10 °C
kW
7.80
8
24.00
27.00
8
-20 °C
kW
4.80
4
17.70
19.0
8
-30 °C
kW
3.00
4
11.0
12.0
8
-40 °C
kW
1.60
4
7.5
7.7
8
Cooling water flow
rate
l/min
26.0
---
29.0
29.0
---
Cooling water tem-
perature
minimum/maximum
°C
10/30
---
10/30
10/30
---
V6
Integral Process Thermostats and High-Temperature Thermostats
171 / 198