Personnel:
n
Operating personnel
1.
Select the [Reset all] menu item.
2.
Select one of the following options:
n
By pressing the [no] key, you return to the higher-level menu
without making any changes.
n
By pressing the [yes] key, the device is reset to factory settings
when you confirm this action with the Enter key.
The complete device is reset to factory settings.
With the menu [Only control parameters] or [Other parameters], only the
selected parameters are reset to factory settings.
Personnel:
n
Operating personnel
1.
Select the [Other parameters] menu item.
2.
Select one of the following options:
n
By pressing the [no] key, you return to the higher-level menu
without making any changes.
n
By pressing the [yes] key, the device is reset to factory settings
when you confirm this action with the Enter key.
The parameters are reset to factory settings.
6.16
Viewing the Device status
In the Device Status menu and its submenus, you can display lots of infor-
mation and data about the device. No settings are possible in the whole
Device Status menu.
1.
Press the Enter key to open the menu.
2.
Select the menu items
Setup
Device Status
.
A submenu opens.
3.
These menu items are available:
n
Error memory,
n
Display device data,
n
Display operating information,
n
Display heating limiter,
n
Display software version,
n
Display hardware version,
n
Display device type and
n
Display serial numbers.
The device has an error memory for storing up to 140 warning, error and
alarm messages that can be used to analyze errors.
Select and reset individual control parame-
ters
Displaying the error memory
V6
Integral Process Thermostats and High-Temperature Thermostats