42 / 54
Alpha
10
Technical data and diagrams
The figures have been determined according to DIN 12876.
Common technical data of Alpha Thermostats
Ambient temperature range
°C
5 – 40
Humidity
Maximum relative humidity 80 % for temperatures up to 31 °C,
decreasing linearly to 50 % relative humidity at 40 °C
Storage temperature range
°C
-50 – 70
Temperature stability at 37 °C
K
±0.05
Display & setting resolution
°C
0.1
Pump type / number of pump steps
--
Pressure pump / 1
Connections for consumers
(accessory)
---
Nipples ½”, 12.7 mm external diameter
Heater power
230 V; 50/60 Hz
115 V; 60 Hz
100 V; 50/60 Hz
kW
kW
kW
1.5
1.15
1.0 (
Least mains fuse protection
230 V~
115 V~
100 V~
A
A
A
12
15
15
Safety Class
---
1/NFL* according to DIN 12876
Protection class
---
Protection class l according to DIN EN 61140; VDE 0140-1
*NFL non-flammable liquids
Product specifications are subject to change without notice!
EU conformity
The device complies with the basic health and safety requirements outline in the Directives listed below.
Machinery Directive 2006/42/EC
EMC Directive 2014/30/EU
LAUDA DR. R. WOBSER GMBH & CO. KG – Pfarrstraße 41/43 - 97922 Lauda-Königshofen -
Germany
The device does not fall under Pressure Equipment Directive 2014/68/EU because the
device is only classified as high as Category 1 and is covered by the Machinery Directive.
Summary of Contents for Alpha
Page 2: ......
Page 4: ......
Page 14: ...14 54 Alpha 09 2020 j 1 Control head with clamp 2 Bath vessel 3 Four mounting feet 1 2 3 ...
Page 16: ...16 54 Alpha 09 2020 j 1 Mains cable 2 Rating label 3 Cooler cable 4 Bath drain nozzle 1 2 3 4 ...
Page 47: ...09 2020 j Alpha 47 54 Cooling characteristics Alpha RA 8 RA 12 and RA 24 230 V 230 V ...
Page 51: ...09 2020 j Alpha 51 54 intentionally left blank ...
Page 52: ...52 54 Alpha 09 2020 j intentionally left blank ...
Page 54: ...54 54 Alpha 08 2020 i ...
Page 55: ......